THE TRADE PAPER OF THE FACTORY FLOOR
Z-1712-2015Class II

Oncentra Brachy radiation therapy planning software designed to analyze and plan radiation treatments in...

Elekta Limited / initiated 2015-05-21 / Terminated

Elekta, Inc. began a recall of Oncentra Brachy radiation therapy planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients... on . FDA classified it as Class II and gives the reason as "Incorrect dose calculation for Regions of Interest (ROIs) defined on a secondary image series". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1712-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Oncentra Brachy radiation therapy planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Incorrect dose calculation for Regions of Interest (ROIs) defined on a secondary image series.
FDA root cause
Other
Action
Important User Notice, 806-03-BTP-001, was distributed to customers on 5/21/2015, A software patch is in development and is expected to be released end of November, 2015. Elekta service will work with customers to implement the patch on affected devices within 6 months of the patch release.
Quantity in commerce
351
Distribution
AL, AZ, AR, CA, CO, CT, DC, FL, GA, HI, IL, IN, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NC, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WV, WI, Argentina, Australia, Bangladesh, Belarus, Botswana, Brazil, Canada, Chile, China, Colombia, Ecuador, Egypt, Guatemala, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kosovo, Libya, Malaysia, Mexico, Morocco, Namibia, Nepal, New Zealand, Oman, Pakistan, Peru, Philippines, Qatar, Russia, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Sri Lanka, Suriname, Taiwan, Thailand, Tobago, Ukraine, Venezuela, Vietnam, Zambia
Product codes and lots
MUJ/892.5050, Software Builds - 3.3, 4.0, 4.1, 4.3

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1712-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1712-2015Class IITerminatedVoluntary: Firm initiated