R3(tm) 28 Mm I.d., 60 Mm O.d. Constrained Liner, Ref 71339160, Qty: (1), Sterile Eo
Smith & Nephew, Inc. began a recall of R3(TM) 28 MM I.D., 60 MM O.D. CONSTRAINED LINER, REF 71339160, QTY: (1), STERILE EO, Product Usage: The R3 Constrained liner Acetabular System is a cemented or... on . FDA classified it as Class II and gives the reason as "The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1738-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- R3(TM) 28 MM I.D., 60 MM O.D. CONSTRAINED LINER, REF 71339160, QTY: (1), STERILE EO, Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained liner is intended for single use only.
- Product code
- KWZ Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer
- Reason for recall
- The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.
- FDA root cause
- Process change control
- Action
- Smith & Nephew sent an Urgent-Product Recall Notification letter dated May 6, 2015. to Sales offices by E-mail and Hospitals via Fed- Ex. The letter identified the affected product, problem and actions to be taken. Customers were asked to quarantine the product and return it. Distributors were asked to notified their customers of the recall and ask them to return the product.
- Quantity in commerce
- 40 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, FL, IL, IN, MI, NE, NY, TN, TX, VA, WI and the countries of Australia, Switzerland, Chile, Canada, and Spain
- Product codes and lots
- Lot numbers: 16HM08087, 14HM08097, 14HM08133, 14HM08141, and 14JM21517
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1735-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1736-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1737-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1738-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1739-2015 | Class II | Terminated | Voluntary: Firm initiated |