Z-1847-2015Class II

Medtronic Navigated Cannulated Taps Individual cannulated taps are packaged in a clear plastic tube with...

Medtronic Navigation, Inc / initiated 2015-05-26 / Terminated / root cause: labeling and UDI

Medtronic Navigation, Inc. began a recall of Medtronic Navigated Cannulated Taps Individual cannulated taps are packaged in a clear plastic tube with protective plastic end caps. Kit/sets contain various... on . FDA classified it as Class II and gives the reason as "Medtronic Navigation is initiating a field correction due to the potential for injury which could occur as a result of cleaning the Medtronic Navigated Cannulated Taps". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1847-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic Navigated Cannulated Taps Individual cannulated taps are packaged in a clear plastic tube with protective plastic end caps. Kit/sets contain various devices including the taps. The kit/set configuration varies dependent on the part number. The devices are assembled and packaged into a shipping carton or into a holding tray which is then placed into a shipping carton. Designed for use in spine surgery to assist in the tapping procedure and the placement of screws during the surgery.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Medtronic Navigation is initiating a field correction due to the potential for injury which could occur as a result of cleaning the Medtronic Navigated Cannulated Taps.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Consignees were notified via Urgent Medical Device Correction Notification on 5/26/15. The letter identified the affected product and described the issue. This field correction does not require the return of the subject products. Revised instructions for use are being provided to the identified consignees.
Quantity in commerce
15,798
Distribution
Worldwide Distribution.
Product codes and lots
Correction includes all lots of products distributed between January 2007 and March 2015.  P/N: 9734240,Nav 6.5mm Tap (Cannulated),9734239, Nav 5.5mm Tap (Cannulated), 9734238,Nav 4.5mm Tap (Cannulated), 9734302,6.5mm Tap (Cannulated),9734300,5.5mm Tap (Cannulated),9734298,4.5mm Tap (Cannulated),9734303,7.5mm Tap (Cannulated),9734301,6.0mm Tap (Cannulated),9734299,5.0mm Tap (Cannulated),9734304,8.5mm Tap (Cannulated),9730944,4.5 mm Cannulated Tap,9730945,5.5 mm Cannulated Tap,9730946,6.5 mm Cannulated Tap,9733182,4.0 mm Cannulated Tap, 1.86P,9733183,4.5 mm Cannulated Tap,9733184,4 mm Cannulated Tap, 1.5P, UCSS,960-347,7.5 Old Round APT (Cannulated Tap-Tip only),960-345,6.5 Old Round APT (Cannulated Tap-Tip only),960-344,5.5 Old Round APT (Cannulated Tap-Tip only) The above referenced cannulated taps are included in the kits and,9731921,KIT 9731921 INSTRUMENT SPINE BASIC,9735101,INST KIT 9735101 TAP ASM 4.5 G5, 9735102,INST KIT 9735102 TAP ASM 5.5 G5,9735103,INST KIT 9735103 TAP ASM 6.5 G5,9735107,INST KIT 9735107 TAP ASM 5.0 G5,9735108,INST KIT 9735108 TAP ASM 6.0 G5,9735109,INST KIT 9735109 TAP ASM 7.5 G5,9735149,INST KIT 9735149 CANNULATED 6.5MM TAP,9735150,INST KIT 9735150 CANNULATED 5.5MM TAP,9735151,INST KIT 9735151 CANNULATED 4.5MM TAP,9735168,INST KIT 9735168 4MM TAP 1.86P,9735169,INST KIT 9735169 4.5MM CANN TAP,9735170,INST KIT 9735170 4MM TAP 1.5P,9734490,INST SET 9734490 LEGACY,9734491,INST SET 9734491 SOLERA STD,9734492,INST SET 9734492 SOLERA DEF UPGRADE,9735469,TAPS 9735469 CD HORIZON SOLERA,9734631,TD SET 9734631 LEGACY STD TAPS/DRIVERS,9734632,TD SET 9734632 SOLERA STD TAPS/DRIVERS,9733370,VERTEX MAX SET, 9733370, NAVIGATED,9734838,KIT 9734838 SOLERA 8.5 CANNULATED TAP,

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1847-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1847-2015Class IITerminatedVoluntary: Firm initiated