Pull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling is...

Synthes GmbH / initiated 2015-05-28 / Terminated / root cause: design

Synthes, Inc. began a recall of Pull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling is placed through the guide sleeve and plate holes to... on . FDA classified it as Class II and gives the reason as "Certain affected parts and lots of the Pull Reduction Devices may have been manufactured to an incorrect hardness specification which could result in intraoperative breakage". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1799-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Pull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling is placed through the guide sleeve and plate holes to pull or push bone fragments relative to the plate. This instrument can be used for: Minor Varus-valgus adjustments (approximately 2 - 4), Translational adjustments, Stabilization of plate bone orientation during insertion of the first screws, alignment of segmental fragments and pre-drilling dense or thick cortical bone before placing a 5.0 mm locking screw.
Product code
FZX Guide, Surgical, Instrument
Reason for recall
Certain affected parts and lots of the Pull Reduction Devices may have been manufactured to an incorrect hardness specification which could result in intraoperative breakage.
Root cause tag
design
FDA root cause
Component design/selection
Action
DePuy Synthes sent an Urgent Notice Urgent notice dated May 28, 2015, to all affected customers to advise them of the defect and patient risks associated with its use. Customers were instructed if they have any of the identified devices, please take the following steps: Call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number. Complete the Verification Section (page 3 of this letter) by checking the appropriate box indicating affected product has been located. Also, please indicate the number of devices found and note the Return Authorization Number. Please include your name, title, address, telephone number and signature in the spaces provided. Return the Verification Section (page 3 of this letter) with the product to: Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132. Send a copy of the completed Verification Section by: Fax: (888) 943-5206 or Scan/email: [email protected] If you DO NOT have the identified product, please take the following steps: Complete the attached Verification Section (page 3 of this letter) by checking the appropriate box indicating that no affected product has been located. Please include your name, title, address, telephone number and signature in the spaces provided. This return documentation acknowledges your receipt of medical device removal information. Return the documents by: Fax: (888) 943-5206 or Scan/email: [email protected] Note: If the Verification Section is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual on page 3 of the notification. Customers with questions were instructed to call 610-719-5450. For questions regarding this recall call 610-719-5000.
Quantity in commerce
3747 US, 19 Foreign
Distribution
Worldwide Distribution-US (nationwide) and internationlly to Canada
Product codes and lots
Part Number: 03.120.023, with lot numbers:  1825699 1825946 1959344 1959346 3052800 3066995 3066996 3066997 3092693 3092737 3139434 3139480 3139483 3139485 3159152 3167363 3172147 3207364 3218030 3218068 3223545 3224270 3224271 3224275 3224581 3224582 3224583 3233445 3288206 3422091 3447606 3469240 3502469 3535235 3556421 3556422 3584316 3642892 3740917 3740919 3740920 3740924 3775480 3775482 3811200 7510644 7510645 7510647 7510648 7542178 7555756 7629268 7699116 7699117 7742370 7742371 7742372 7778806 7778807 7821741 7878176 7995396 8016667 8028575 8068778 8070887 8070891 8195690 8210669 8217914 8408715

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1799-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1799-2015Class IITerminatedVoluntary: Firm initiated