Z-1831-2015Class II

SERFAS 90 degree Energy Probe, Part Number 279-350-101; SERFAS Energy Probes are indicated for arthroscopic...

Stryker Corp. / initiated 2015-06-03 / Terminated / root cause: design

Stryker Endoscopy began a recall of SERFAS 90 degree Energy Probe, Part Number 279-350-101; SERFAS Energy Probes are indicated for arthroscopic procedures of the knee, shoulder, ankle, hip, elbow... on . FDA classified it as Class II and gives the reason as "Stryker Endoscopy is recalling all non expired SERFAS 90 degree Energy Probes due to reports of fragments of the probe breaking off into the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1831-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
SERFAS 90 degree Energy Probe, Part Number 279-350-101; SERFAS Energy Probes are indicated for arthroscopic procedures of the knee, shoulder, ankle, hip, elbow and wrist. Specifically, the probes are used for resection, ablation and coagulation of soft tissue, as well as the hemostasis of blood vessels.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Stryker Endoscopy is recalling all non expired SERFAS 90 degree Energy Probes due to reports of fragments of the probe breaking off into the patient.
Root cause tag
design
FDA root cause
Device Design
Action
Stryker sent an Urgent Medical Device Recall letter dated June 5, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to do the following: 1. Inform individuals within their organization who need to be aware of this device removal. 2. Review inventory of lots of part number 279-350-101 and determine if you have the affected product (all non-expired devices) in stock. Response is required. 3. If no product is found, complete acknowledgement form located on the Stryker Endoscopy recall website endorecall.stryker.com by logging in using the account number and zip code on this letter. 4. If you do have product, segregate the product and call Stryker customer service at 1-800-624-4422 Option 3 to arrange for product return and issuance of credit. For questions regarding this recall call 408-754-2664.
Quantity in commerce
22,063 devices
Distribution
Worldwide Distribution - US (nationwide) and Internationally to US Argentina, Australia, Bolivia, Brazil, Chile, China, Colombia, Hong Kong, India, Italy, Japan, Republic of Korea, Malaysia, Netherlands, Poland, Romania, Singapore, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Product codes and lots
Part number 279-350-101; All non expired product; lot numbers 13128AE2 through 14337AE2.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1831-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1831-2015Class IITerminatedVoluntary: Firm initiated