Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5...
Volcano Corp began a recall of Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5 - 804200-002 and 804200-014; s5x - 807300-001; CORE... on . FDA classified it as Class II and gives the reason as "A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR)...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2043-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Volcano Corporation
- Product
- Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5 - 804200-002 and 804200-014; s5x - 807300-001; CORE MOBILE - 400-0100.01. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.
- Product code
- IYO System, Imaging, Pulsed Echo, Ultrasonic
- Reason for recall
- A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Volcano and Philips sent an Advisory Notice letter dated September 23, 2013 to all impacted customers. The letter identified the affected product, problem and actions to be taken. For question contact your Volcano and Philips representative.
- Quantity in commerce
- 227 total units affected by the issue - all models
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Belgium, United Kingdom, Denmark, Netherlands, Sweden, Iceland, France, Saudi Arabia, Australia, Philippines and Thailand.
- Product codes and lots
- s5 Part number 804200-002 and 804200-014; s5x Part number 807300-001; CORE MOBILE Part Number 400-0100.01. All software versions
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2043-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2044-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2045-2015 | Class II | Terminated | Voluntary: Firm initiated |