Z-2043-2015Class II

Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5...

Volcano Corporation / initiated 2013-09-23 / Terminated / root cause: manufacturing process

Volcano Corp began a recall of Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5 - 804200-002 and 804200-014; s5x - 807300-001; CORE... on . FDA classified it as Class II and gives the reason as "A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR)...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2043-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Volcano Corporation
Product
Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5 - 804200-002 and 804200-014; s5x - 807300-001; CORE MOBILE - 400-0100.01. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.
Product code
IYO System, Imaging, Pulsed Echo, Ultrasonic
Reason for recall
A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Volcano and Philips sent an Advisory Notice letter dated September 23, 2013 to all impacted customers. The letter identified the affected product, problem and actions to be taken. For question contact your Volcano and Philips representative.
Quantity in commerce
227 total units affected by the issue - all models
Distribution
Worldwide Distribution - US Nationwide and the countries of Belgium, United Kingdom, Denmark, Netherlands, Sweden, Iceland, France, Saudi Arabia, Australia, Philippines and Thailand.
Product codes and lots
s5 Part number 804200-002 and 804200-014; s5x Part number 807300-001; CORE MOBILE Part Number 400-0100.01. All software versions

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2043-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2043-2015Class IITerminatedVoluntary: Firm initiated
Z-2044-2015Class IITerminatedVoluntary: Firm initiated
Z-2045-2015Class IITerminatedVoluntary: Firm initiated