Z-2366-2015Class II

Oasis Coil Extension Cable used with Hitachi Oasis MRI System

FUJIFILM Healthcare Americas Corporation / initiated 2014-04-25 / Terminated / root cause: manufacturing process

Hitachi Medical Systems America Inc began a recall of Oasis Coil Extension Cable used with Hitachi Oasis MRI System Product Usage: The Oasis MR system is an imaging device and is intended to provide the physician... on . FDA classified it as Class II and gives the reason as "The firm received a complaint stating the Coil Extension Cable accessory would overheat and be a burn hazard if touched". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2366-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Oasis Coil Extension Cable used with Hitachi Oasis MRI System Product Usage: The Oasis MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The Oasis system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination. The Coil Extension Cable that is the subject of the removal action is an accessory that is used in conjunction with certain anatomy coils to aid in patient positioning.
Product code
MOS Coil, Magnetic Resonance, Specialty
Reason for recall
The firm received a complaint stating the Coil Extension Cable accessory would overheat and be a burn hazard if touched.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Hitachi sent an Urgent: Device Removal Notice letter date April 24, 2014 to their customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers to discontinue using the Coil Extension Cable and that a Field Service Engineer will schedule a visit to remove the cable. For questions call 800-800-3106 ext 3720.
Quantity in commerce
173 units
Distribution
US Nationwide in the states: AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI & WY and the country of Mexico.
Product codes and lots
Serial #'s: M001to M108, M110 to M175, M951 & M952.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2366-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2366-2015Class IITerminatedVoluntary: Firm initiated