510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2209-2015Class II

InjectionPlasty 1.0 Bone Void Filler; an injectable and biocompatible calcium phosphate bone void filler...

Skeletal Kinetics, LLC / initiated 2015-05-29 / Terminated

Skeletal Kinetics, Llc began a recall of InjectionPlasty 1.0 Bone Void Filler; an injectable and biocompatible calcium phosphate bone void filler. Sterilized using irradiation. The single-use... on . FDA classified it as Class II and gives the reason as "Reports of venous uptake, DVT, pulmonary embolism in patients who were implanted with Injection Plasty Bone Void Filler". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2209-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Skeletal Kinetics, LLC
Product
InjectionPlasty 1.0 Bone Void Filler; an injectable and biocompatible calcium phosphate bone void filler. Sterilized using irradiation. The single-use InjectionPlasty 1.0 Kit contains the necessary components for mixing and delivery of the bone void filler. The InjectionPlasty 1.0 sterile kit contains: Calcium Phosphate Powder, Dilute Sodium Silicate Liquid, a Mixing System, Targeting K-Wires and a Delivery Needle (Cannula). Model number 10801210. InjectionPlasty 1.0 is indicated to fill bony voids or gaps of the skeletal system
Product code
MQV Filler, Bone Void, Calcium Compound
Reason for recall
Reports of venous uptake, DVT, pulmonary embolism in patients who were implanted with Injection Plasty Bone Void Filler.
FDA root cause
No Marketing Application
Action
Skeletal Kinetics sent letters to distributors by overnight mail with instructions to forward a Dear Doctor notification letter on May 14, 2015. Second letters dated May 29, 2015, advised that the firm is removing the product from the marker. Distributor letters asked that the Dear Doctor letter be forward within 24 hours of receipt of the letter. All Injection Plasty kits in their possession and those with customers should be retrieved and returned by Fed Ex. Dear Doctor letters advised that their distribution agent has been requested to return all unused kits. Questions may be directed to 408-366-5002 of [email protected].
Quantity in commerce
67 kits
Distribution
US Distribution including Puerto Rico.
Product codes and lots
Model number: 10801210: Lot Numbers Expiration Dates  13022501 8/31/2015 13052305 11/30/2015 13120901 5/31/2016 14051901 11/30/2016 14100101 3/31/2017 15042701 10/13/2017

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2209-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2209-2015Class IITerminatedVoluntary: Firm initiated