Z-1947-2015Class II

Etest Cefotaxime CT 256 WW and Etest Cefotaxime CT 256 US. in vitro diagnostic

bioMerieux, Inc. / initiated 2015-06-12 / Terminated / root cause: manufacturing process

BioMerieux SA began a recall of Etest Cefotaxime CT 256 WW and Etest Cefotaxime CT 256 US. in vitro diagnostic on . FDA classified it as Class II and gives the reason as "QC results were out of range resulting in false resistant strains". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1947-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
Etest Cefotaxime CT 256 WW and Etest Cefotaxime CT 256 US. in vitro diagnostic.
Product code
JWY Manual Antimicrobial Susceptibility Test Systems
Reason for recall
QC results were out of range resulting in false resistant strains.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
BioMerieux sent an Urgent Product Removal Notices dated June 12, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Required actions: Please distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties that may use this product, including others to whom you may have transferred our product. Discontinue using and discard any remaining inventory of Etest CEFOTAXIME CT 256 lots reported in table 1. Discuss any concerns you may have regarding previously reported results with your Laboratory Medical Director to determine the appropriate course of action. Complete and return the Acknowledgement Form in Attachment A by Fax to confirm receipt of this notice and to request product replacement. bioMrieux is committed to providing our customers with the highest quality product possible. We sincerely apologize for any inconvenience that this may have caused you. If you require additional assistance or have any questions, please contact your local bioMrieux Clinical Customer Service organization at (800) 682-2666.
Quantity in commerce
674
Distribution
Worldwide Distribution - USA (nationwide) to the states of AL, AR, CA, CO, FL, GA, IL, IN, KY, MD, MI, MN, MO, NE, NY, NC, OH, OK, OR, PA, TN, TX, UT, VA, WA, WV and WI., and to the countries of: Austria, Australia, Canada, Switzerland, Germany, Spain, France, United Kingdom, Greece, Hong Kong, India, Italy, Japan, Netherlands, Poland, Sweden, Thailand, Turkey, Saudi Arabia, Philippines, Slovenia, Kuwait, Brazil, Chile, Czech Republic, Algeria, South Africa, Kosovo, Costa Rica, Belarus, Guam, Honduras and Vietnam.
Product codes and lots
1002774320, 1002377890, 1002590470, 1002775330, 1002379610

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1947-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1947-2015Class IITerminatedVoluntary: Firm initiated