Z-2785-2015Class II

Radiation Therapy Treatment Planning System software RayStation is a software system designed for treatment...

RaySearch Laboratories AB (publ) / initiated 2015-05-29 / Terminated / root cause: software

RAYSEARCH LABORATORIES AB began a recall of Radiation Therapy Treatment Planning System software RayStation is a software system designed for treatment planning and analysis of radiation therapy. The... on . FDA classified it as Class II and gives the reason as "An error caused by a floating point precision problem tends to cause a problem with the display of isodose lines for dose and dose difference and color tables for dose, dose difference and PET images". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2785-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
RaySearch Laboratories AB (publ)
Product
Radiation Therapy Treatment Planning System software RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
An error caused by a floating point precision problem tends to cause a problem with the display of isodose lines for dose and dose difference and color tables for dose, dose difference and PET images
Root cause tag
software
FDA root cause
Software design
Action
The firm, RaySearch Laboratories, distributed a "Field Safety Notice, Medical Device Correction #7106" notice dated May 29, 2015,and reply forms to their customers via email. The notice describes the product, problem and actions to be taken. The Customers were provided a workaround as a temporary fix. Customers were also instructed to educate all users, inspect your product and identify all installed units, confirm you have read and understood the notice by contacting Quality and Regulatory Affairs Manager; and complete and return the reply form via email to [email protected] or by fax to 888-501-7195. The solution will be resolved in the next version of RayStation, scheduled for release in June 2015. Customers were advised to share the notice to all those who need to be aware within the organizations. Customers can contact Quality and Regulatory Affairs Manager at +46 8 510 530 12 or via email at [email protected].
Quantity in commerce
20 devices
Distribution
Worldwide Distribution: US (nationwide) to state of WA; and countries of: France, Sweden, Austria, Spain, Germany, Italy, UK, New Zealand.
Product codes and lots
software build numbers are 4.7.0.15 or 4.7.1.10

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2785-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2785-2015Class IITerminatedVoluntary: Firm initiated