Z-1933-2015Class II

SOMATOM Emotion 16; produce cross-sectional images of the body by computer reconstruction of x-ray...

Siemens Medical Solutions USA, Inc. / initiated 2015-06-04 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Emotion 16; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at... on . FDA classified it as Class II and gives the reason as "A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1933-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Emotion 16; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site.
Root cause tag
design
FDA root cause
Device Design
Action
A safety advisory notice, dated June 4, 2015, was sent to direct accounts explaining the problem and potential risk to patients and operators. Siemens engineers will begin replacing the bolts in May 2015.
Quantity in commerce
40 total
Distribution
Nationwide Distribution.
Product codes and lots
model numbers: 10165977, with serial numbers: 32310 32254 32322 32317 32303 32295 32327 32309 32319 32237 32326 32259 32323 32257 32298 32315 32268 32244 32252 32293 32349 32313 32334 32301 32339 32277 32292 32281 32273 32320 32324 32346 32270

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1933-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1933-2015Class IITerminatedVoluntary: Firm initiated
Z-1934-2015Class IITerminatedVoluntary: Firm initiated