510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2091-2015Class II

Helical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to...

Synthes (USA) Products LLC / initiated 2015-06-10 / Terminated / root cause: manufacturing process

Synthes (USA) Products LLC began a recall of Helical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent... on . FDA classified it as Class II and gives the reason as "The height of one of the three guiding pins of the Helical Blade Inserter for the TFNA is oversized. The Helical Blade Inserter is therefore unable to pass through the Blade/Screw Guide Sleeve (part 03.037.017)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2091-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes (USA) Products LLC
Product
Helical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element insertion. Manual orthopedic surgical instrument.
Product code
HSB Rod, Fixation, Intramedullary And Accessories
Reason for recall
The height of one of the three guiding pins of the Helical Blade Inserter for the TFNA is oversized. The Helical Blade Inserter is therefore unable to pass through the Blade/Screw Guide Sleeve (part 03.037.017).
Root cause tag
manufacturing process
FDA root cause
Process control
Action
A recall notification letter, dated June 10, 2015, was sent to sales consultants and Synthes distribution sites requesting return of affected devices.
Quantity in commerce
32
Distribution
Nationwide Distribution.
Product codes and lots
Part Number Lot Numbers 03.037.024 T102759

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2091-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2091-2015Class IITerminatedVoluntary: Firm initiated