510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1934-2015Class II

SOMATOM Emotion 6; produce cross-sectional images of the body by computer reconstruction of x-ray...

Siemens Medical Solutions USA, Inc. / initiated 2015-06-04 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Emotion 6; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at... on . FDA classified it as Class II and gives the reason as "A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1934-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Emotion 6; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site.
Root cause tag
design
FDA root cause
Device Design
Action
A safety advisory notice, dated June 4, 2015, was sent to direct accounts explaining the problem and potential risk to patients and operators. Siemens engineers will begin replacing the bolts in May 2015.
Quantity in commerce
40 total
Distribution
Nationwide Distribution.
Product codes and lots
model numbers: 10165888 with serial numbers: 29161 29185 29186 29165 29176 29181 29198

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1934-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1933-2015Class IITerminatedVoluntary: Firm initiated
Z-1934-2015Class IITerminatedVoluntary: Firm initiated