SOMATOM Emotion 6; produce cross-sectional images of the body by computer reconstruction of x-ray...
Siemens Medical Solutions USA, Inc began a recall of SOMATOM Emotion 6; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at... on . FDA classified it as Class II and gives the reason as "A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1934-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- SOMATOM Emotion 6; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- A safety advisory notice, dated June 4, 2015, was sent to direct accounts explaining the problem and potential risk to patients and operators. Siemens engineers will begin replacing the bolts in May 2015.
- Quantity in commerce
- 40 total
- Distribution
- Nationwide Distribution.
- Product codes and lots
- model numbers: 10165888 with serial numbers: 29161 29185 29186 29165 29176 29181 29198
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1933-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1934-2015 | Class II | Terminated | Voluntary: Firm initiated |