Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1955-2015Class II

JustRight 5mm Stapler, JR-ST25-2.0 Stapler is packaged in cases of 6, placed in a tray and sealed with a...

Bolder Surgical, LLC / initiated 2015-06-11 / Terminated / root cause: manufacturing process

Justright Surgical, LLC began a recall of JustRight 5mm Stapler, JR-ST25-2.0 Stapler is packaged in cases of 6, placed in a tray and sealed with a Tyvek lid. Product Usage: The JustRight 5mm Stapler is... on . FDA classified it as Class II and gives the reason as "JustRight Surgical is recalling various lots of JustRight 5mm Stapler product because they may deliver incomplete staple formation and incomplete tissue closure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1955-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bolder Surgical, LLC
Product
JustRight 5mm Stapler, JR-ST25-2.0 Stapler is packaged in cases of 6, placed in a tray and sealed with a Tyvek lid. Product Usage: The JustRight 5mm Stapler is intended for use in abdominal, gynecologic, pediatric, and thoracic surgery for the resection, transection, and creation of anastomosis.
Product code
GDW Staple, Implantable
Reason for recall
JustRight Surgical is recalling various lots of JustRight 5mm Stapler product because they may deliver incomplete staple formation and incomplete tissue closure.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
JustRight Surgical sent an Urgent Medical Device Recall letter dated June 12, 2015 to aaffected customers and distributor via an email. The letter identified the affected product, problem and actions to be taken. Custers were instructed Review, complete, and return the enclosed Acknowledgement Form directly to Rep, your Sales Representative or email to [email protected] or fax to (720) 287-7135. The following assistance is available: " For questions related to the issue or product replacement, please contact your local sales representative, Rep, or contact JustRight Surgical Customer Service at (866) 683-1743, Monday through Friday from 7:30 AM to 5:00 PM MST. " For questions related to product pick-up by JustRight Surgical, please also contact your local sales representative or Customer Service (866) 683-1743. Return instructions are provided on the Acknowledgement Form. " For questions related to the Acknowledgement Form and its return, please contact the JustRight Surgical Director of Quality Systems at (720) 287-7146.
Quantity in commerce
822
Distribution
Distribution - US Nationwide and Ireland.
Product codes and lots
Model Number JR-ST25-2.0 Lot numbers 75ED0116, 75JD1404, 75KD0308, 75KD1216, 75LD0508, 75AE0709, 75AE2601, 75BE0506, 75BE2315, 75CE1630, 75CE2606, 75DE0102, and 75DE0804 (UK distribution).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1955-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1955-2015Class IITerminatedVoluntary: Firm initiated
Z-1956-2015Class IITerminatedVoluntary: Firm initiated