Z-0205-2016Class II

Arthroscopic Energy 50 Probe with Suction, Catalog number: AES-50S Peoduct Usage: For resection, ablation...

ConMed Corporation / initiated 2015-06-17 / Terminated / root cause: software

ConMed Corporation began a recall of Arthroscopic Energy 50 Probe with Suction, Catalog number: AES-50S Peoduct Usage: For resection, ablation, and coagulation of soft tissue and hemostasis of... on . FDA classified it as Class II and gives the reason as "Probe programming error. The incorrect program can generate more heat at the surgical site, resulting in an elevated probability of tissue damage". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0205-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConMed Corporation
Product
Arthroscopic Energy 50 Probe with Suction, Catalog number: AES-50S Peoduct Usage: For resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic surgical procedures.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Probe programming error. The incorrect program can generate more heat at the surgical site, resulting in an elevated probability of tissue damage.
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
On June 17, 2015 Con Med Corporation distributed Urgent Device Recall Notification Letters and Recall Reply forms to their customers via courier service. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine their inventory for any of the affected lot numbers referenced on the recall notice. If any of the recalled product are in stock, customers should discontinue the use or distribution of these items. If product was further distributed, customers should notify those individuals of this recall action. Customers with questions can call Gregory Connell at 727-399-5276 or email [email protected]. Customers should also complete and return the Recall Reply Form by mail to the address listed on the recall notice.
Quantity in commerce
276 devices
Distribution
Worldwide - US distribution in the states of FL, OH, and the country of Spain.
Product codes and lots
1412161, 1410201, 1410271, 1412121, and 1412081

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0205-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0205-2016Class IITerminatedVoluntary: Firm initiated