510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2060-2015Class II

ZNN Antegrade Femoral Nail (ZNN AF). Orthopedic internal fixation device

Zimmer GmbH / initiated 2015-06-17 / Terminated / root cause: manufacturing process

Zimmer, Inc. began a recall of ZNN Antegrade Femoral Nail (ZNN AF). Orthopedic internal fixation device on . FDA classified it as Class II and gives the reason as "A single distributed ZNN Greater Trochanter Femoral Nail may have been dented. Potential for fatigue of implant prior to sufficient fracture healing, which may result in need for implant removal and revision of fracture...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2060-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ZNN Antegrade Femoral Nail (ZNN AF). Orthopedic internal fixation device.
Product code
HSB Rod, Fixation, Intramedullary And Accessories
Reason for recall
A single distributed ZNN Greater Trochanter Femoral Nail may have been dented. Potential for fatigue of implant prior to sufficient fracture healing, which may result in need for implant removal and revision of fracture fixation to assure proper healing and avoid fracture malunion or nonunion.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 6/17/2015, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Managers with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761. Hours of operation are Monday through Friday, 8 a.m. through 5 p.m. EST.
Quantity in commerce
8
Distribution
Worldwide distribution. US states of NC and AK; Taiwan, Japan, and Germany.
Product codes and lots
Part 47-2492-320-10; lot 62763724

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2060-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2060-2015Class IITerminatedVoluntary: Firm initiated