Z-2102-2015Class II

Stryker SelectCore Variable Throw Biopsy Device, 14G, 10CM

Stryker Corp. / initiated 2015-06-02 / Terminated / root cause: manufacturing process

Stryker Instruments Div. of Stryker Corporation began a recall of Stryker SelectCore Variable Throw Biopsy Device, 14G, 10CM Product Usage: The Stryker SelectCore Variable Throw Biopsy Device is used in obtaining biopsy... on . FDA classified it as Class II and gives the reason as "Stryker Instruments is recalling the Stryker SelectCore Variable Throw Biopsy Device due to the potential for for the stylet to eject from the SelectCore device during actuation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2102-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker SelectCore Variable Throw Biopsy Device, 14G, 10CM Product Usage: The Stryker SelectCore Variable Throw Biopsy Device is used in obtaining biopsy samples from soft tissues such as liver, breast, kidney, prostate, spleen, lymph nodes, and various soft tissue tumors. For breast biopsy this product is for diagnosis onlynot therapeutic use.. These devices are sterile, single use devices.
Product code
KNW Instrument, Biopsy
Reason for recall
Stryker Instruments is recalling the Stryker SelectCore Variable Throw Biopsy Device due to the potential for for the stylet to eject from the SelectCore device during actuation.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Stryker distributed Customer Notification Letters via Fed Ex overnight to be delivered on 06/10/2015. Customers are asked to Immediately review this Recall Notification. 2. Immediately check all stock areas and/or operating room storage and quarantine any affected product found. 3. Complete the enclosed Business Reply Form (BRF) to confirm receipt of this notification and identify how many, if any, affected items are currently in your inventory. Please complete and return the BRF even if you do not have any affected product on hand. Note: Your signature on the BRF indicates that you received and understand this Notification and have followed the instructions in this Notification. 4. If you have further distributed this product, please forward this letter and the attached Business Reply Form (BRF) to all affected locations. Please indicate each location on the BRF. 5. Fax (866-521-2762) or email ([email protected]) the completed Business Reply Forms; and return the recalled SelectCore devices(s) to Stryker.
Quantity in commerce
55 units
Distribution
US Nationwide Distribution in the states of: TX, NJ and MS.
Product codes and lots
Stryker Model Number: 0904-410-000; Lot# 61406001

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2102-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2102-2015Class IITerminatedVoluntary: Firm initiated