Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number...
Pega Medical Inc. began a recall of Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number SCF-MLD265 and SCF-MLD273 on . FDA classified it as Class II and gives the reason as "SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1848-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Orthopediatrics Canada ULC
- Product
- Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number SCF-MLD265 and SCF-MLD273.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation.
- FDA root cause
- Reprocessing Controls
- Action
- Pega Medical sent an Urgent Field Safety-Recall letter via email on June 15, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Advise on action to be taken by the customer: Identify the device with the same catalogue number and lot number Return the device to Pega Medical Inc. to the address shown below using the FedEx account. For further questions please call the firm at 450-688-5144 x. 242.
- Quantity in commerce
- 4 units
- Distribution
- US Distribution to the states of : FL and TN.
- Product codes and lots
- Part Number: SCF-MLD265, LOT#: 130703-01. Part Number: SCF-MLD273, LOT#: 130703-02.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1848-2015 | Class II | Terminated | Voluntary: Firm initiated |