Z-1985-2015Class II

GE Healthcare Optima CL 323i. Angiographic X-ray systems used in generating fluoroscopic and rotational...

GE Healthcare / initiated 2015-05-29 / Terminated / root cause: manufacturing process

GE Healthcare began a recall of GE Healthcare Optima CL 323i. Angiographic X-ray systems used in generating fluoroscopic and rotational images of human anatomy for diagnostic, interventional... on . FDA classified it as Class II and gives the reason as "GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1985-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare Optima CL 323i. Angiographic X-ray systems used in generating fluoroscopic and rotational images of human anatomy for diagnostic, interventional and image-guided surgical procedures.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery).
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter on 5/29/2015- GEHC Ref#12231- dated May 29, 2015 to their Consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and contact Information. The customers were instructed that you may continue to use the system and avoid rapid direction changes with the gantry motions and prevent potential collisions with other equipment. If the failure occurs during interventional examination, please ensure that you have established procedures for handling patients in case of the loss of gantry motion as stated in your product labeling. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for this correction. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
7
Distribution
Worldwide Distribution: US (nationwide) including states of: CA, DE, FL, GA, IL, IN, KS, MS, NV, OH, OK, OR, PA, TN, TX, VA, WA and WI; and countries of: ARGENTINA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAPAN, RUSSIA, UNITED KINGDOM, and VIETNAM.
Product codes and lots
Serial Number System ID Not Available GON4271838 Not Available 229890CL Not Available 361788CL323 Not Available 806358INNOVA 0C2V069A4104R1 082416310033 0C2V069A4202R1 082416200021 0C2V069A4203R1 082416050013

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1985-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1973-2015Class IITerminatedVoluntary: Firm initiated
Z-1974-2015Class IITerminatedVoluntary: Firm initiated
Z-1975-2015Class IITerminatedVoluntary: Firm initiated
Z-1976-2015Class IITerminatedVoluntary: Firm initiated
Z-1977-2015Class IITerminatedVoluntary: Firm initiated
Z-1978-2015Class IITerminatedVoluntary: Firm initiated
Z-1979-2015Class IITerminatedVoluntary: Firm initiated
Z-1980-2015Class IITerminatedVoluntary: Firm initiated
Z-1981-2015Class IITerminatedVoluntary: Firm initiated
Z-1982-2015Class IITerminatedVoluntary: Firm initiated
Z-1983-2015Class IITerminatedVoluntary: Firm initiated
Z-1984-2015Class IITerminatedVoluntary: Firm initiated
Z-1985-2015Class IITerminatedVoluntary: Firm initiated