Z-2063-2015Class II

MOSAIQ Oncology Information System MOSAIQ is an oncology information system used to manage workflows for...

Elekta Limited / initiated 2015-07-01 / Terminated / root cause: software

Elekta, Inc. began a recall of MOSAIQ Oncology Information System MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports... on . FDA classified it as Class II and gives the reason as "A problem exists in MOSAIQ resulting in the incorrect field size being sent to the treatment machine for stereotactic plans using cones". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2063-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
MOSAIQ Oncology Information System MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
A problem exists in MOSAIQ resulting in the incorrect field size being sent to the treatment machine for stereotactic plans using cones.
Root cause tag
software
FDA root cause
Software design
Action
Elekta sent an Urgent Important Field Safety Notice 371-01-MSQ-003 on July 1, 2015, to all affected customers. The notice informed users of the specific product and version numbers affected and that the problem only exists if all the following criteria is met - (1) Running Mosaiq 2.41; (2) with at least one linac equipped with an Agility 160 MLC head (Versa HD, Infinity, Axesse (EOS) or Synergy); and (3) who are treating with Stereotactic Cones. The notice also advises users of the workaround to avoid the problem. The customers were also instructed to return the Important Field Safety Notice Acknowledgment form. For questions regarding this recall call 770-670-2422.
Quantity in commerce
74 Total (45 US)
Distribution
Worldwide Distribution - US including AZ, CA, GA, ID, IL, MO, NJ, NY, NC, OR, SC, TN, WV, and Internationally to Austria, Brazil, Bulgaria, Canada, Colombia, Germany, Israel, Italy, Japan, Mexico, Russia, Taiwan, and Venezuela.
Product codes and lots
Software Version 2.41 and higher

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2063-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2063-2015Class IITerminatedVoluntary: Firm initiated