Z-2271-2015Class II

Fabius MRI Anesthesia Machine; Fabius MRI is an inhalation anesthesia machine for use in MRI environments in...

Draeger, Inc. / initiated 2015-06-22 / Terminated / root cause: labeling and UDI

Draeger Medical, Inc. began a recall of Fabius MRI Anesthesia Machine; Fabius MRI is an inhalation anesthesia machine for use in MRI environments in operating, induction and recovery rooms on . FDA classified it as Class II and gives the reason as "the Fabius MRI or parts of the system were attracted by the magnetic field of the MRI. If the Fabius MRI is positioned too close to the MRI, its parts can be loosened as a result of the magnetic force of the MRI". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2271-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger, Inc.
Product
Fabius MRI Anesthesia Machine; Fabius MRI is an inhalation anesthesia machine for use in MRI environments in operating, induction and recovery rooms.
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
the Fabius MRI or parts of the system were attracted by the magnetic field of the MRI. If the Fabius MRI is positioned too close to the MRI, its parts can be loosened as a result of the magnetic force of the MRI.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
Draeger Medical sent an Urgent Medical Device Recall Letter, dated June 2015, and Instructions for Use Supplement (one supplement for each Fabius MRI Anesthesia Machine at the facility) to end users. The letter identified the affected product, problem and actions to be taken. For questions regarding the recall letter contact Michael Kelhart at 1-800-543-5047 (press 1 at the prompt, then 2, then 32349). For questions regarding the operation and/or servicing of the Drger Fabius MRI anesthesia machine contact DrgerService Technical Support at 1-800-543-5047 (press 4 at the prompt).
Quantity in commerce
373
Distribution
US Nationwide Distribution
Product codes and lots
Catalog Number(s): 8607300

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2271-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2271-2015Class IITerminatedVoluntary: Firm initiated