Sorin Stockert Heater-Cooler 3T, 230 V Temperature control for extracorporeal perfusion of durations up to 6...
Sorin Group USA, Inc. began a recall of Sorin Stockert Heater-Cooler 3T, 230 V Temperature control for extracorporeal perfusion of durations up to 6 hours on . FDA classified it as Class II and gives the reason as "Potential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2076-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- Sorin Stockert Heater-Cooler 3T, 230 V Temperature control for extracorporeal perfusion of durations up to 6 hours.
- Product code
- DWC Controller, Temperature, Cardiopulmonary Bypass
- Reason for recall
- Potential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Sorin Group issued a Field Safety Notice dated June 15, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory and identify any affected devices. For each unit customers were instructed to determine if the device has been maintained according to the Instructions for Use. If yes, customers should strictly adhere to the new Instructions for Use. Customers were also provide with a Response form to confirm they received, read and understood the Field Notice. Customers were instructed to return the completed form to assist in monitoring the effectiveness of the communication. For technical support customers should call 1-800-221-7943, ext 6355. For questions regarding this recall call 303-467-6306.
- Quantity in commerce
- 2837
- Distribution
- Worldwide Distribution - US (nationwide) and Internationally to AL, AT, BE, BG, CH, CY, CZ, DE, DK, EE, ES, FI, FR, GB, GR, HR, HU, IE, IT, LT, LU, LV, MC, MK, MT, NL, NO, PL, PT, RO, RS, SE, SK, AE, AR, AU, AZ, BD, BH, BR, BY, CA, CL, CN, CO, CR, DZ, EC, EG, ET, GE, GY, HK, ID, IL, IN, IQ, IR, JO, JP, KR, KW, KZ, LB, LK, LY, MA, MN, MU, MX, MY, NG, NP, NZ, OM, PA, PE, PH, PK, PR, PS, QA, RE, RU, SA, SG, SV, SY, TH, TN, TR, TT, TW, UA, VN, ZA.
- Product codes and lots
- Product code 16-02-80 Serial number 16S10027-16S15641
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2076-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2077-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2078-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2079-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2080-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2081-2015 | Class II | Terminated | Voluntary: Firm initiated |