Contour Next Control Solution for Contour Next Blood Glucose Test System. Liquid quality control solution...

Medrad, Inc. / initiated 2015-06-22 / Terminated / root cause: design

Bayer Healthcare, LLC began a recall of Contour Next Control Solution for Contour Next Blood Glucose Test System. Liquid quality control solution. Plastic bottle, 2.5mL. Indicated for use with fresh... on . FDA classified it as Class II and gives the reason as "If the control solution bottle is not thoroughly shaken prior to use, the control test result may be higher than the control range and may not be recognized and marked as a control test by the blood glucose meter. Also...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2482-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medrad, Inc.
Product
Contour Next Control Solution for Contour Next Blood Glucose Test System. Liquid quality control solution. Plastic bottle, 2.5mL. Indicated for use with fresh fingertip capillary whole blood samples.
Product code
JJX Single (Specified) Analyte Controls (Assayed And Unassayed)
Reason for recall
If the control solution bottle is not thoroughly shaken prior to use, the control test result may be higher than the control range and may not be recognized and marked as a control test by the blood glucose meter. Also, the control solution, used as a quality control test in blood glucose meter systems, may read outside the specified range too high or too low.
Root cause tag
design
FDA root cause
Device Design
Action
On 6/22/2015, Bayer Healthcare began sending URGENT - MEDICAL DEVICE FIELD CORRECTIVE ACTION notifications to affected Health Care Providers, Pharmacies, and Users. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers are instructed to always check the expiration, not to use and discard of expired control solutions. Also, only CONTOUR NEXT control solution should be used with the CONTOUR NEXT test strips. The control solution bottle should be shaken about 15 times before each use. The letter instructed customers to refer to their meter's user guide for further testing instructions. If test results outside the control range are received, another test should be performed following the instructions provided in the letter. If the problem persists, customers should contact Bayer Diabetes Support. Bayer will replace the control solution or provide new control ranges for the test strips, depending on the situation. The firm has implemented improvements to provide more consistent control solution product.
Quantity in commerce
5 million bottles of control solution
Distribution
Worldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Croatia, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Slovenia, Spain, Sweden, Switzerland, UK, Albania, Azerbaijan, Bulgaria, Czech Republic, Estonia, Hungary, Korea, Latvia, Lithuania, Philippines, Romania, Serbia, Slovakia, and Thailand.
Product codes and lots
SKU# 7314 (Level 2) and SKU# 7315 (Level 1) All Lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2482-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2482-2015Class IITerminatedVoluntary: Firm initiated