510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2323-2015Class I

ReSTOR intraocular lens and ReSTOR Toric intraocular lens. The Alcon AcrySof IQ ReSTOR and IQ ReSTOR Toric...

Alcon Laboratories / initiated 2015-04-15 / Terminated / root cause: manufacturing process

Alcon Research, Ltd. began a recall of ReSTOR intraocular lens and ReSTOR Toric intraocular lens. The Alcon AcrySof IQ ReSTOR and IQ ReSTOR Toric intraocular lenses are posterior chamber lenses... on . FDA classified it as Class I and gives the reason as "There are reports of post - operative inflammation and/or toxic anterior segment syndrome (TASS) after cataract surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2323-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alcon Laboratories
Product
ReSTOR intraocular lens and ReSTOR Toric intraocular lens. The Alcon AcrySof IQ ReSTOR and IQ ReSTOR Toric intraocular lenses are posterior chamber lenses intended as optical implants designed to replace the human crystalline lens in adult patients following cataract surgery. The lenses subject to this correction are manufactured specifically for Japan.
Product code
HQL Intraocular Lens
Reason for recall
There are reports of post - operative inflammation and/or toxic anterior segment syndrome (TASS) after cataract surgery.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Alcon Research sent recall notification letters dated April 16, 2015, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to stop using the affected product and place on hold until their sales representatives visit their site for withdrawal. Follow up communication activities included personal contact via phone and site visits. For questions regarding this recall call 817-293-0450.
Quantity in commerce
45,391
Distribution
Distribution Japan, No US distribution.
Product codes and lots
Model Numbers: SN6AD1, SN6AD3, SV25T0, MN6AD1,SND1T3, SND1T4, SND1T5 and SND1T6 and all lot numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2323-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2323-2015Class ITerminatedVoluntary: Firm initiated