RadiForce RX650, 6MP Color LCD Monitor, Part Code No. 0FTD1957A5, 0FTD195B5, UPC 690592037203. This product...
Eizo Corporation began a recall of RadiForce RX650, 6MP Color LCD Monitor, Part Code No. 0FTD1957A5, 0FTD195B5, UPC 690592037203. This product is intended to be used in displaying and viewing... on . FDA classified it as Class II and gives the reason as "EIZO is recalling the RadiForce LCD Monitors could display abnormally while connecting to particular graphics boards with DisplayPort". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2088-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Eizo Corporation
- Product
- RadiForce RX650, 6MP Color LCD Monitor, Part Code No. 0FTD1957A5, 0FTD195B5, UPC 690592037203. This product is intended to be used in displaying and viewing digital images for review and analysis by trained medical practitioners. It does not support the display of mammography images for diagnosis.
- Product code
- PGY Display, Diagnostic Radiology
- Reason for recall
- EIZO is recalling the RadiForce LCD Monitors could display abnormally while connecting to particular graphics boards with DisplayPort.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- EIZO sent An Urgent Field Safety Notice dated July 7, 2015, to all customers who purchased the RadiForce RX440, RX650, RX850 and GX540. The notice informs customers that the RadiForce monitors could display abornormally in 6-bit (64 gradations) instead of normal 8-bit (256 gradations) or 10-bit (1024 gradations) while connecting to particular graphics boards with DisplayPort. The notice provides the customers with the actions to be taken and contact information (1-800-800-5202, M-F, 8am-4pm, PST, [email protected], 5710 Warland Drive, Cypress, CA 90630).
- Quantity in commerce
- 3 units
- Distribution
- Nationwide Distribution including OH, FL, IL, TX, OH, NY, AZ, and NJ.
- Product codes and lots
- 20326034, 20618064, 20671064
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2087-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2088-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2089-2015 | Class II | Terminated | Voluntary: Firm initiated |