BD Vacutainer SST Plus Blood Collection Tubes, 5.0 mL, 13 x 100 mm, Sterile. Provides a means of collecting...

Becton Dickinson & Co / initiated 2015-06-23 / Terminated / root cause: manufacturing process

Becton Dickinson & Company began a recall of BD Vacutainer SST Plus Blood Collection Tubes, 5.0 mL, 13 x 100 mm, Sterile. Provides a means of collecting, transporting, separating, and processing blood in... on . FDA classified it as Class III and gives the reason as "Some of the tubes were manufactured with stoppers that did not meet current manufacturing specifications. Stoppers were manufactured with a reduced cycle cure time". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2434-2015
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Vacutainer SST Plus Blood Collection Tubes, 5.0 mL, 13 x 100 mm, Sterile. Provides a means of collecting, transporting, separating, and processing blood in a closed tube.
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
Some of the tubes were manufactured with stoppers that did not meet current manufacturing specifications. Stoppers were manufactured with a reduced cycle cure time.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Becton Dickinson notified their affected customers of this recall by sending them an Urgent Product Recall Letter dated June 23, 2015 via UPS 2nd Day.
Quantity in commerce
238, 000 (Lot #5133959) and 243,000 (Lot #5133957)
Distribution
Distributed in FL, IL, MA, MI, MO, PA and TN.
Product codes and lots
REF #367986, Lot #'s 5133957 and 5133959

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2434-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2434-2015Class IIITerminatedVoluntary: Firm initiated