510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2265-2015Class II

IPS Empress Universal Glaze Spray 270 mL, Product Code/Model/Catalog Number 609432AN, Rx only. Porcelain...

Ivoclar Vivadent, Inc. / initiated 2015-07-09 / Terminated / root cause: manufacturing process

Ivoclar Vivadent, Inc. began a recall of IPS Empress Universal Glaze Spray 270 mL, Product Code/Model/Catalog Number 609432AN, Rx only. Porcelain glaze spray used in the fabrication of dental... on . FDA classified it as Class II and gives the reason as "Complaints were received of coarse particles in the glazed surface of IPS e.max Ceram restorations. A basic component of the glass contains a higher number of large grain sizes than usual". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2265-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ivoclar Vivadent, Inc.
Product
IPS Empress Universal Glaze Spray 270 mL, Product Code/Model/Catalog Number 609432AN, Rx only. Porcelain glaze spray used in the fabrication of dental restorations
Product code
EIH Powder, Porcelain
Reason for recall
Complaints were received of coarse particles in the glazed surface of IPS e.max Ceram restorations. A basic component of the glass contains a higher number of large grain sizes than usual.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Ivoclar Vivadent sent an URGENT MEDICAL DEVICE RECALL Letter (dated 7/8/2015) and Response Form to affected customers via Certified Mail Return Receipt Request. The letter identified the affected product, problem and actions to be taken. Customers were instructed to notify dentists of the recall and request return of product. A notification letter was sent to dentists on 7/24/2015 requesting they destroy the product on hand (if any) and report back to Ivoclar Vivodent. For questions in the United States contact US Headquarters Ivoclar Vivadent Customer Services at 1-800-533-6825.
Quantity in commerce
Domestic: 27 units
Distribution
US Distribution in the states of: PA, IN, IA and SC.
Product codes and lots
Batch U15124

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2265-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2265-2015Class IITerminatedVoluntary: Firm initiated