Z-2484-2015Class I

OmniPod Insulin Management System. Only the OmniPod device is affected. Catalog Number(s): POD-ZXP420 (US)...

Insulet Corporation / initiated 2015-07-13 / Terminated / root cause: manufacturing process

Insulet Corporation began a recall of OmniPod Insulin Management System. Only the OmniPod device is affected. Catalog Number(s): POD-ZXP420 (US). Intended for subcutaneous delivery of insulin at... on . FDA classified it as Class I and gives the reason as "OmniPods (Pods) have a higher rate of failure causing: Cannula fails to deploy/retract causing insulin to be pumped or the audible alarm is displayed on the PDM and pod will not deliver insulin". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2484-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Insulet Corporation
Product
OmniPod Insulin Management System. Only the OmniPod device is affected. Catalog Number(s): POD-ZXP420 (US). Intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.
Product code
LZG Pump, Infusion, Insulin
Reason for recall
OmniPods (Pods) have a higher rate of failure causing: Cannula fails to deploy/retract causing insulin to be pumped or the audible alarm is displayed on the PDM and pod will not deliver insulin
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Insulet Corporation, sent an "URGENT: Voluntary Device Recall" letter dated 7/13/2015 to each consignee/customer via Email notification (where possible) on 7/13/15 as well as Federal Express. Customers not responding to the email or Federal Express will receive follow up phone calls. The letter described the product, problem and actions to e taken. The customers were instructed to do the following: If you have Pods from any of these lots, set them aside, we will replace them at no charge; arrange for return and replacement by choosing one of the following methods: 1. Complete the online form at https://www.myomnipod.com/podinfo (fastest method). 2. Return the enclosed reply card by mail or fax to 1-855-407-3729. 3. Contact Insulet Customer Care by calling 1-855-407-3729 (any time day or night). If you do not have any unused Pods form these lots, reply using one of the methods above. If you have any question, contact Acting VP RA/QA at 978-600-7000. EXPANDED RECALL: Insulet Corp. issued on 11/2/15 the URGENT: Field Safety Notification via Email notification and Federal Express. Letter describes problem of that certain lots of the pod's needle failed to deploy or there is a delay in the deployment of the needle. Customers not responding to the email or Federal Express will receive additional mailing and/or follow up phone calls. Call Customer Care at 1-855-407-3729 if you have any questions regarding this Field Safety Notification
Quantity in commerce
24829 boxes Expanded: 26,230.9 boxes
Distribution
Worldwide Distribution: US (Nationwide) and country of: Switzerland.
Product codes and lots
Lot Numbers: L40806, L40811, L40895, L40976, L41014, L41025, L41067, L41162, L41171, L41197, L41198, L41250  Expanded Recall: Lot Codes: L41880, L41881, L41892, L41895,L41897,L41898, L41899,L41900, L41901, L41902, L41903, L41904, L41905, L41906, L41907

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2484-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2484-2015Class ITerminatedVoluntary: Firm initiated
Z-2485-2015Class ITerminatedVoluntary: Firm initiated