Z-2376-2015Class II

PSW 1203.U/1, PSW 1301.U, PSW 1307.U, PSW 1403.U, and PSW 1501.U Pacemaker Programmer software. Software...

BIOTRONIK AG / initiated 2015-06-30 / Terminated / root cause: software

BIOTRONIK, Inc. began a recall of PSW 1203.U/1, PSW 1301.U, PSW 1307.U, PSW 1403.U, and PSW 1501.U Pacemaker Programmer software. Software approved for use with BIOTRONIK device programmers... on . FDA classified it as Class II and gives the reason as "Ventricular packing: LV software programming versions for BIOTRONIK CRT-P and CRT-D devices are recalled because they are not approved for use in the US". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2376-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BIOTRONIK AG
Product
PSW 1203.U/1, PSW 1301.U, PSW 1307.U, PSW 1403.U, and PSW 1501.U Pacemaker Programmer software. Software approved for use with BIOTRONIK device programmers (trade names: Renamic and ICS 3000). Allows physicians to program devices to pace exclusively in the left ventricle.
Product code
rce
Reason for recall
Ventricular packing: LV software programming versions for BIOTRONIK CRT-P and CRT-D devices are recalled because they are not approved for use in the US.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
BIOTRONIK sent the Important Product Information letter and the Acknowledgement of Important Product Information Form, dated June 20, 2015, to affected physicians with the following recommendations: -Review programming settings for your patients with CRT-P or CRT-D devices. -If you have a patient with a device programmed to Ventricular pacing: LV, arrange to have the patients device reprogrammed to an appropriate approved mode (i.e. "Ventricular pacing: RV" or "Ventricular pacing: BiV"). -Physicians are requested to sign the Acknowledgement of Important Product Information Form and immediately fax to 800-913-6993 or email to [email protected]. BIOTRONIK'S representative will update your programmer with approved 1502.U programmer software. Following this update, the Ventricular pacing: LV feature will no longer be available. However, a device previously programmed with "Ventricular pacing: LV will remain programmed in this mode unless and until the pacing mode setting is changed. If you have any questions concerning this Important Product Information, please contact the BIOTRONIK Technical Services Department: Telephone: 1-(800) 284-6689 (24-hour service)
Quantity in commerce
768 units (software)
Distribution
Nationwide Distribution.
Product codes and lots
Affected Product Part Numbers: 394229, 400335, 403222, 406797, 408849

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2376-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2376-2015Class IITerminatedVoluntary: Firm initiated