Z-2257-2015Class II

Computed Tomography X-ray Systems Ingenuity Core. Intended to produce cross-sectional images of the body by...

Philips North America LLC / initiated 2015-02-05 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Computed Tomography X-ray Systems Ingenuity Core. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data... on . FDA classified it as Class II and gives the reason as "5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2257-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Computed Tomography X-ray Systems Ingenuity Core. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value incorrectly remains constant for some heart rate change conditions. DoseRight incorrectly enabled when the Exam Card specified disabling DoseRight. Tube heat predict
Root cause tag
software
FDA root cause
Software change control
Action
On 2/5/2015 the firm sent Customer Information Letters to their customers.
Quantity in commerce
82 units
Distribution
Worldwide Distribution-US (nationwide) including the states of AR, AZ, CA, CT, DE, FL, GA, IA, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VT, WV & WY, and to the following countries: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cambodia, Canada, China, Colombia, Denmark, Finland, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Italy, Japan, Latvia, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Norway, Panama, Philippines, Russia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan & Viet Nam.
Product codes and lots
Model #728321; Serial #s: 31001, 52007, 52008, 52010, 52012, 52013, 52018, 52020, 52021, 52024, 52026, 52030, 52031, 52032, 52033, 52034, 52035, 52036, 52037, 52038, 52039, 52042, 52043, 52046, 52052, 52053, 52054, 52055, 52057, 52058, 310001, 310002, 310003, 310009, 310017, 310052, 310055, 310059, 310064, 310069, 310070, 310071, 310081, 310087, 310088, 310096, 310111, 310113, 310119, 310121, 310129, 310133, 310134, 310141, 310143, 310144, 310149, 310151, 310157, 310160, 310170, 333002, 333009, 333014, 333015, 333030, 333031, 333037, 333042, 333043, 333044, 333045, 333046, 333051, 333052, 333053, 333055, 333056, 333057, 333059, 333060 & 333061.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2257-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2254-2015Class IITerminatedVoluntary: Firm initiated
Z-2255-2015Class IITerminatedVoluntary: Firm initiated
Z-2256-2015Class IITerminatedVoluntary: Firm initiated
Z-2257-2015Class IITerminatedVoluntary: Firm initiated
Z-2258-2015Class IITerminatedVoluntary: Firm initiated
Z-2259-2015Class IITerminatedVoluntary: Firm initiated