i-Stat PT/INR Cartridges List number 03P89-24 The cartridges are packaged in sealed pouches, one per pouch...
Abbott Point Of Care Inc. began a recall of i-Stat PT/INR Cartridges List number 03P89-24 The cartridges are packaged in sealed pouches, one per pouch. There are 24 cartridges packed in a box. a... on . FDA classified it as Class II and gives the reason as "Certain lots of PT/INR cartridges may generate a higher than typical number of PT/INR star outs potentially delaying the generation of patient results for PT/INR". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2361-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- i-Stat PT/INR Cartridges List number 03P89-24 The cartridges are packaged in sealed pouches, one per pouch. There are 24 cartridges packed in a box. a prothrombin time test useful in monitoring patients receiving oral anticoagulation therapy such as coumadin or warfarin.
- Product code
- GJS Test, Time, Prothrombin
- Reason for recall
- Certain lots of PT/INR cartridges may generate a higher than typical number of PT/INR star outs potentially delaying the generation of patient results for PT/INR.
- FDA root cause
- Under Investigation by firm
- Action
- Abbott Point of Care sent a recall letter dated June 15, 2015 via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return the enclosed Business Reply Card (BRC). A Proof of Destruction From (FG-4533 in included with letter and is required to be returned to APOC as part of the credit process. Customers with questions were instructed to contact APOC Technical Support at 800-366-8020, Option 1 or their APOC representative.
- Quantity in commerce
- 423,168 cartridges
- Distribution
- Worldwide Distribution - US (nationwide) and Internationally to China, Lebanon, Costa Rica, Singapore, Brazil, Jordan, Russia, Greece, Austria, India, Nigeria, United Kingdom, Korea, Saudi Arabia, Morocco, Slovenia, Portugal, Tunisia, Kazakhstn, Qatar, U.A.E. Israel, Taiwan, Mexico, Argentina, Poland, Italy, Hong Kong, Kuwait, Switzerland, Turkey, Ireland, Chile, Germany, and Australia.
- Product codes and lots
- All lots from S15007 to S15026 inclusive and All lots from T15037A to T15076A inclusive.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2361-2015 | Class II | Terminated | Voluntary: Firm initiated |