Z-2329-2015Class I

Puritan Bennett 980 Ventilator System, Universal (with neonatal functionality enabled) and Neonatal Models...

Medtronic, Inc. / initiated 2015-07-16 / Terminated / root cause: software

Covidien LP (formerly Nellcor Puritan Bennett Inc.) began a recall of Puritan Bennett 980 Ventilator System, Universal (with neonatal functionality enabled) and Neonatal Models. Designed for use on Neonatal (NICU) through Adult... on . FDA classified it as Class I and gives the reason as "Reports in which tidal volumes reaching patients were lower than set tidal volumes in neonatal Volume Control Plus (VC+) Mode with active humidification. This situation may potentially lead to respiratory compromise if...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2329-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Puritan Bennett 980 Ventilator System, Universal (with neonatal functionality enabled) and Neonatal Models. Designed for use on Neonatal (NICU) through Adult patient populations who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb).
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Reports in which tidal volumes reaching patients were lower than set tidal volumes in neonatal Volume Control Plus (VC+) Mode with active humidification. This situation may potentially lead to respiratory compromise if not recognized.
Root cause tag
software
FDA root cause
Software design
Action
Consignees will be notified via letter on July 17, 2015. No products will be returned, the reconfiguration of affected ventilators will occur at customer sites.
Quantity in commerce
657 units
Distribution
Worldwide Distribution-US (nationwide) and the countries of Canada, United Arab Emirates, Argentina, Australia, Bahrain, Brazil, Chile, Colombia, Costa Rica, Egypt, Ireland, Israel, Jordan, Japan, Kuwait, Lebanon, Morocco, Mexico, Panama, Peru, Qatar, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Turkey, and Uruguay.
Product codes and lots
PB980 Ventilator (980xxxxxxxx): 980U1ENDIUU, 980U1ENDIUUS, 980U3ENDIUU, 980U3ENDIUUS, 980N1ENDIUU, 980N1ENDIUUS, 980N3ENDIUU, 980N3ENDIUUS

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2329-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2329-2015Class ITerminatedVoluntary: Firm initiated