uHead, Ulnar Implant, Standard Stem, Size 1, Sterile, Rx Only. Intended for replacement of the distal...
Stryker Howmedica Osteonics Corp. began a recall of uHead, Ulnar Implant, Standard Stem, Size 1, Sterile, Rx Only. Intended for replacement of the distal radioulnar joint following ulnar head resection... on . FDA classified it as Class II and gives the reason as "Stryker is recalling rHead, uHead, Sigmoid Notch, Remotion, Radio Capitellum because packaging integrity (sterile barrier) of the packaging type KIT I may potentially be compromised by transportation". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2427-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- uHead, Ulnar Implant, Standard Stem, Size 1, Sterile, Rx Only. Intended for replacement of the distal radioulnar joint following ulnar head resection arthroplasty.
- Product code
- KXE Prosthesis, Wrist, Hemi-, Ulnar
- Reason for recall
- Stryker is recalling rHead, uHead, Sigmoid Notch, Remotion, Radio Capitellum because packaging integrity (sterile barrier) of the packaging type KIT I may potentially be compromised by transportation.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- Branches/Agencies were notified by Stryker of this action by e-mails on June 24, 2015 and an Urgent Medical Device Removal Letter/Acknowledgement Response Form dated June 25, 2015 was sent to the attention of the Risk Manager.
- Quantity in commerce
- 16992 total
- Distribution
- Worldwide Distribution.
- Product codes and lots
- Part Number: UHA-S1 (Ulnar Stem Implant #1 Sterile Packed) - Lot #'s 17584, 17585, 22217301, 22217302, 23470801, 19066-S2, 19417-S2, 19418-S2, 1973-S2, 19732-S2, 19785-S2, 19815-S2 and $23470801; Part Number: UHA-S120 (Recon Ulnar Stem Implant #1, 20mm Extension) - Lot #'s 16725, 16726, 18577, 18757, 18758, 18759, 18867, 23124001, 19056-S2, 19786-S2, 19816-S2, 19928-S2 and 19929-S2; Part Number: UHA-S2 (Ulnar Stem Implant #2 Sterile Packed) - Lot #'s 16714, 16715, 18760, 22760101, 22760102, 23124101, 23124102, 23470901, 19669-S2, 19817-S2, 19866-S2 and R23470901; Part Number: UHA-S220 (Recon Ulnar Stem Implant #2, 20mm Extention) - Lot #'s 16720, 16721, 18415, 18761, 18762, 18763, 18764, 18765, 18868, 23124201, 19122-S2, 19787-S2, 19818-S2, 19930-S2 and 19932-S2; Part Number: UHA-S3 (Ulnar Stem Implant #3 Sterile Packed) - Lot #'s 16716, 16717, 22760201, 22760202, 22760203, 23124301, 23124302, 19788-S2, 19789-S2, 19818-S2, 19820-S2, 19821-S2, 19867-S2 and 19868-S2; Part Number: UHA-S320 (Recon Ulnar Stem Implant #3, 20mm Extention) - Lot #'s16722, 16723, 18869, 19065-S2, 19157-S2, 19733-S2, 19790-S2 and 19933-S2; Part Number: UHA-S4 (Ulnar Stem Implant #4 Sterile Packed) - Lot #'s 16718, 16719, 18766, 18865, 18866, 19156-S2, 19670-S2, 19671-S2, 19672-S2, 19734-S2, 19822-S2 and 19869-S2; Part Number: UHA-S420 (Recon Ulnar Stem Implant #4, 20mm Extention) - Lot #'s 16724, 18578, 18767, 18768, 18870, 18871, 22760301, 19072-S2, 19100-S2, 19158-S2, 19791-S2, 19823-S2, 19934-S2 and 19935-S2
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2426-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2427-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2428-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2429-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2430-2015 | Class II | Terminated | Voluntary: Firm initiated |