510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2200-2015Class II

Pinnacle3 Software Version 9.0, 9.2, 9.4 and 9.6, Model Numbers 453560446041, 459800091001, 459800220161...

Philips North America LLC / initiated 2014-07-31 / Terminated / root cause: software

Philips Electronics North America Corporation began a recall of Pinnacle3 Software Version 9.0, 9.2, 9.4 and 9.6, Model Numbers 453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451. UPDATED... on . FDA classified it as Class II and gives the reason as "Philips, Pinnacle Radiation Treatment Planning System version 8.0h, 8.0k, 8.0m, 8.0n, 9 0, 9 2 9 4, 9 6 is being recalled because the dose may be inconsistent with the density of a density-overridden ROl". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2200-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Pinnacle3 Software Version 9.0, 9.2, 9.4 and 9.6, Model Numbers 453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451. UPDATED: Pinnacle3 Software Version 8.0h, 8.0k, 8.0m, 8.0n.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Philips, Pinnacle Radiation Treatment Planning System version 8.0h, 8.0k, 8.0m, 8.0n, 9 0, 9 2 9 4, 9 6 is being recalled because the dose may be inconsistent with the density of a density-overridden ROl.
Root cause tag
software
FDA root cause
Software design
Action
Consignees were sent on 7/31/2014 a Philips "Urgent Field Safety Notice" dated 2014-July 20. The letter described the problem, product involved in the recall and the action that should be taken by the customers. For questions contact Customer Care Solutions Center at 1-800-722-9377. UPDATE: A second Philips letter dated 2 July 2015 was sent in August 6, 2015. This letter updated the recall to incorporate the additional products being recalled.
Quantity in commerce
1642
Distribution
Worldwide Distribution-US: All states in the US including DC and PR. OUS: Australia, Austria, Bangladesh, Belgium, Brazil, China, Denmark, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands,New Zealand, Oman, Philippines, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Viet Nam. UPDATED: Canada, Bulgaria, Egypt, Hungary, Jordan, Luxembourg, Poland, Reunion, Romania, Russia, Sauth Arabia, Turkey.
Product codes and lots
453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451  UPDATED:  453560446051 [A]  459800096361 [Al 453560446061 IA]  459800096541 [A] 453560446091 [A]  459800096741 [A] 453560446101 [A]  459800096921 [A] 453560446111 [A]  459800097111 [A] 453560446131 [A]  459800097291 [A] 453560446141 [A]  459800097471 [A] 453560446161 [A]  459800097651 [B] 453560446181 [A]  459800097831 [B] 453560446201 [A]  459800098171 [B] 459800089221 [A]  459800098351 [A] 459800091451 [A]  459800109261 [A] 459800138101 [BJ 459800189431 [A) 459800220131 [B] 459800239171 [A] 459800233271 [8] 459800233281 [A] 459800233291 [B] 459800233301 [A] 459800233311 [8] 459800233321 [B] 459800233331 [B] 459800233341 [A] 459800233351 [A] 459800233361 [A] 459800234101 [A] 459800234111 [A] 459800234121 [A] 459800318391 [Al 459800405751 [A] 459800417961 [A] 459800249651 [8] 459800249721 [A] 459800249851 [A] 459800249931 [A] 459800250001 [Al 459800250081 [A] 459800250151 [A] 459800250221 [A] 459800250291 [A] 459800250361 [A] 459800250431 [A] 459800250501 [A] 459800250571 [A] 459800338961 [B] 459800340181 [A] 459800340201 [A] 459800341171 [A] 459800341191 [A] 459800341481 [A] 459800341501 [A] 459800345001 [A] 459800345021 [A] 459800345041 [A] 459800345061 [A] 459800345081 [A] 459800345101 [A] 459800454251 [A] 459800455341 [A] 459800493591 [A] 459800494011 [A] 459800554831 [A] 459800554881 [A] 453560435761 453560434881 453560454791 453560461331 453560461341

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2200-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2200-2015Class IITerminatedVoluntary: Firm initiated