Z-2317-2015Class I

Beacon Tip Torcon NB Advantage Catheter. Angiographic catheter. For use in angiographic procedures as a...

Cook Incorporated / initiated 2015-07-02 / Terminated / root cause: design

Cook Inc. began a recall of Beacon Tip Torcon NB Advantage Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can help... on . FDA classified it as Class I and gives the reason as "Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2317-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Beacon Tip Torcon NB Advantage Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can help physicians diagnose occlusion or stenosis. The catheters are designed for percutaneous introduction into the vascular system over an appropriately sized wire guide.
Product code
DQO Catheter, Intravascular, Diagnostic
Reason for recall
Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result.
Root cause tag
design
FDA root cause
Component design/selection
Action
Cook Medical sent an Urgent Medical Device Recall letter dated July 2, 2015, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification and instructions for returning the affected product. All customers were notified via courier. Customers with questions were instructed to contact Cook Medical Customer Relations at (800) 457-4500 or (812) 339-2235. Firm issued press release 8/3/2015.
Quantity in commerce
451550
Distribution
Worldwide Distribution - US (nationwide including Puerto Rico and US Virgin Islands), Algeria, Antigua, Armenia, Austria, Australia, Bahamas, Belgium, Bangladesh, Bulgaria, Bahrain, Brazil, Brunei Darussalam, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Ghana, Great Britain, Greece Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kuwait, Kazakhstan, Republic of Korea, Kuwait, Lebanon, Lithuania, Luxemburg, Latvia, Libya, Macau, Malaysia, Malta, Martinique, Mexico, Mongolia, Morocco, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Sri Lanka, Sweden, Switzerland, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Product codes and lots
All Lots of following Product numbers: HNBR4.0-35-100-P-NS-0, HNBR4.0-35-100-P-NS-C1, HNBR4.0-35-100-P-NS-C2, HNBR4.0-35-100-P-NS-CANADA-041480, HNBR4.0-35-100-P-NS-DAV, HNBR4.0-35-100-P-NS-H1, HNBR4.0-35-100-P-NS-JB1,  HNBR4.0-35-100-P-NS-JB2, HNBR4.0-35-100-P-NS-JB3,  HNBR4.0-35-100-P-NS-JIM,  HNBR4.0-35-100-P-NS-KMP,  HNBR4.0-35-100-P-NS-MPA, HNBR4.0-35-100-P-NS-O1,  HNBR4.0-35-100-P-NS-SIM1,  HNBR4.0-35-100-P-NS-SIM2,  HNBR4.0-35-100-P-NS-SIM3, HNBR4.0-35-100-P-NS-VERT,  HNBR4.0-35-100-P-NS-VTK,  HNBR4.0-35-100-P-NS-WNBG,  HNBR4.0-35-125-P-NS-0, HNBR4.0-35-125-P-NS-DAV,  HNBR4.0-35-125-P-NS-H1,  HNBR4.0-35-125-P-NS-JB1,  HNBR4.0-35-125-P-NS-JB2, HNBR4.0-35-125-P-NS-MPA,  HNBR4.0-35-125-P-NS-SIM2,  HNBR4.0-35-125-P-NS-VERT,  HNBR4.0-35-125-P-NS-VTK, HNBR4.0-35-125-P-NS-WNBG,  HNBR4.0-35-40-P-NS-KMP,  HNBR4.0-35-65-P-NS-0,  HNBR4.0-35-65-P-NS-C1, HNBR4.0-35-65-P-NS-C2,  HNBR4.0-35-65-P-NS-C3,  HNBR4.0-35-65-P-NS-DAV,  HNBR4.0-35-65-P-NS-H1,  HNBR4.0-35-65-P-NS-JB1, HNBR4.0-35-65-P-NS-KMP,  HNBR4.0-35-65-P-NS-MPA,  HNBR4.0-35-65-P-NS-RC1,  HNBR4.0-35-65-P-NS-RC2, HNBR4.0-35-65-P-NS-RDC,  HNBR4.0-35-65-P-NS-RIM,  HNBR4.0-35-65-P-NS-SIM1,  HNBR4.0-35-65-P-NS-SIM2, HNBR4.0-35-80-P-NS-C1,  HNBR4.0-35-80-P-NS-C2,  HNBR4.0-35-80-P-NS-JL2,  HNBR4.0-35-80-P-NS-JL2.5, HNBR4.0-35-80-P-NS-JL3,  HNBR4.0-35-80-P-NS-JR1,  HNBR4.0-35-80-P-NS-JR2,  HNBR4.0-35-80-P-NS-JR3.5, HNBR4.0-35-80-P-NS-RH,  HNBR4.0-35-80-P-NS-RLG,  HNBR4.0-35-80-P-NS-VS,  HNBR4.0-35-80-P-NS-VS1, HNBR4.0-35-80-P-NS-VS2,  HNBR4.0-35-80-P-NS-VS3,  HNBR4.0-NT-100-P-NS-H1,  HNBR4.0-NT-80-P-NS-RC1, HNBR4.0-NT-80-P-NS-RIM,  HNBR4.0-NT-80-P-NS-RLG, HNBR4.1-35-100-P-NS-0,  HNBR4.1-35-100-P-NS-C1, HNBR4.1-35-100-P-NS-C2,  HNBR4.1-35-100-P-NS-DAV,  HNBR4.1-35-100-P-NS-H1,  HNBR4.1-35-100-P-NS-JB1, HNBR4.1-35-100-P-NS-JB2,  HNBR4.1-35-100-P-NS-JB3,  HNBR4.1-35-100-P-NS-KMP,  HNBR4.1-35-100-P-NS-MPA, HNBR4.1-35-100-P-NS-O1,  HNBR4.1-35-100-P-NS-SIM1,  HNBR4.1-35-100-P-NS-SIM2,  HNBR4.1-35-100-P-NS-SIM3, HNBR4.1-35-100-P-NS-VERT,  HNBR4.1-35-100-P-NS-WNBG,  HNBR4.1-35-110-P-NS-JB1,  HNBR4.1-35-110-P-NS-JB2, HNBR4.1-35-110-P-NS-WNBG,  HNBR4.1-35-120-P-NS-JB2,  HNBR4.1-35-130-P-NS-DAV-T,  HNBR4.1-35-40-P-NS-KMP, HNBR4.1-35-65-P-NS-C1,  HNBR4.1-35-65-P-NS-C2,  HNBR4.1-35-65-P-NS-DAV,  HNBR4.1-35-65-P-NS-H1,  HNBR4.1-35-65-P-NS-JB1, HNBR4.1-35-65-P-NS-KMP,  HNBR4.1-35-65-P-NS-LEV1,  HNBR4.1-35-65-P-NS-MPA,  HNBR4.1-35-65-P-NS-RC1, HNBR4.1-35-65-P-NS-RC2,  HNBR4.1-35-65-P-NS-RDC,  HNBR4.1-35-65-P-NS-RIM,  HNBR4.1-35-65-P-NS-SHK1.0, HNBR4.1-35-65-P-NS-SIM1,  HNBR4.1-35-75-P-NS-CHG-D,  HNBR4.1-35-80-P-NS-C1,  HNBR4.1-35-80-P-NS-C2, HNBR4.1-35-80-P-NS-JL2,  HNBR4.1-35-80-P-NS-JL2.5,  HNBR4.1-35-80-P-NS-JL3,  HNBR4.1-35-80-P-NS-JR1, HNBR4.1-35-80-P-NS-JR2,  HNBR4.1-35-80-P-NS-JR3.5,  HNBR4.1-35-80-P-NS-RC1,  HNBR4.1-35-80-P-NS-RH, HNBR4.1-35-80-P-NS-RLG,  HNBR4.1-35-80-P-NS-VS,  HNBR4.1-35-80-P-NS-VS2,  HNBR4.1-NT-100-P-NS-H1, HNBR4.1-NT-80-P-NS-C1,  HNBR4.1-NT-80-P-NS-RC1,  HNBR4.1-NT-80-P-NS-RH,  HNBR4.1-NT-80-P-NS-RIM, HNBR4.1-NT-80-P-NS-RLG,  HNBR5.0-35-65-P-NS-RUC, HNBR5.0-35-90-P-NS-RUC

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2317-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2317-2015Class ITerminatedVoluntary: Firm initiated
Z-2318-2015Class ITerminatedVoluntary: Firm initiated
Z-2319-2015Class ITerminatedVoluntary: Firm initiated