510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2778-2015Class II

Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5, ED-250XL5, ED-450XT5...

FUJIFILM Healthcare Americas Corporation / initiated 2015-06-30 / Terminated / root cause: design

Fujifilm Medical Systems U.S.A., Inc. began a recall of Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5, ED-250XL5, ED-450XT5, and ED-450XL5 Endoscope Operation Manuals on . FDA classified it as Class II and gives the reason as "The suction button on the ED-530XT endoscope may become lodged (stuck) in the active position following the use of x-ray contrast agents during ERCP procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2778-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5, ED-250XL5, ED-450XT5, and ED-450XL5 Endoscope Operation Manuals.
Product code
FDT Duodenoscope And Accessories, Flexible/Rigid
Reason for recall
The suction button on the ED-530XT endoscope may become lodged (stuck) in the active position following the use of x-ray contrast agents during ERCP procedures.
Root cause tag
design
FDA root cause
Device Design
Action
FujiFilm sent a Voluntary Field Correction/Verification Form dated June 30, 2015 to their affected customers.
Quantity in commerce
256 units
Distribution
Distributed in the US and Canada.
Product codes and lots
ED-530XT - Part #202B1237697D; ED-250XT5 and ED-250XL5 - Part #202B1260196A; ED-450XT5 - Part #202B1260211A and ED-450XL5 - Part #202B1237681

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2778-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2778-2015Class IITerminatedVoluntary: Firm initiated