Z-2206-2015Class II

American Science & Engineering Inc,Gemini Cabinet X-ray system

American Science and Engineering, Inc. / initiated 2015-05-22 / Terminated / root cause: supplier

American Science & Engineering, Inc. began a recall of American Science & Engineering Inc,Gemini Cabinet X-ray system on . FDA classified it as Class II and gives the reason as "During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the...". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-2206-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
American Science and Engineering, Inc.
Product
American Science & Engineering Inc,Gemini Cabinet X-ray system.
Product code
RCG Cabinet X-Ray, Security (Includes Baggage X-Ray)
Reason for recall
During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the operator's console and on both ends of the tunnel) are not illuminated.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
American Science Engineering, Inc (AS&E) is developing a fix for this problem that includes updating a software configuration on the X-ray source. You will be notified by field service as soon as this update is available. Please inform all operators of the system about this communication, so they are aware of the need to shut down and re-start the system should this fault occur. We appreciate your understanding. if you have any questions, please feel free to contact AS&E via e-mail at [email protected]. If you are under a service contract, you can also contact your customer service representative. For further questions please call (978) 262-8700.
Quantity in commerce
125 systems
Distribution
Worldwide Distribution - US Distribution to the states of : NY, FL, TN, PA, NJ, DE, ME, TX, CT, MS, NC, MD, MO, VA, DC, NV and GA., and to the countries of : Saudi Arabia, Indonesia, Canada, Netherlands, Hong Kong, Lebanon and Kazakhstan.
Product codes and lots
Gemini systems

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2206-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2206-2015Class IITerminatedVoluntary: Firm initiated