510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2425-2015Class II

Synthes Tibial Nail Wrap Holds 11 Nails and Synthes Intramedullary Nail Wrap Holds 10 Nails; the Nail Wrap is...

Synthes (USA) Products LLC / initiated 2015-07-01 / Terminated / root cause: sterility

Synthes (USA) Products LLC began a recall of Synthes Tibial Nail Wrap Holds 11 Nails and Synthes Intramedullary Nail Wrap Holds 10 Nails; the Nail Wrap is intended to be used as a protective sheath... on . FDA classified it as Class II and gives the reason as "The wrap is for single use for nail sterilization but does not have sufficient testing for the device to be considered a multi-use item". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-2425-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes (USA) Products LLC
Product
Synthes Tibial Nail Wrap Holds 11 Nails and Synthes Intramedullary Nail Wrap Holds 10 Nails; the Nail Wrap is intended to be used as a protective sheath specifically for Universal Femoral Nails, Universal Tibial Nails, Unreamed Tibial Nail, and the Stainless Steel Tibial Nail during steam sterilization by the hospital.
Product code
FSM Tray, Surgical, Instrument
Reason for recall
The wrap is for single use for nail sterilization but does not have sufficient testing for the device to be considered a multi-use item.
Root cause tag
sterility
FDA root cause
Under Investigation by firm
Action
DePuy Synthes sent an Urgent Notice Medical Device Recall letter dated July 8, 2015, to all affected customers.The letter identified the product the problem and the action needed to be taken by the customer. ACTIONS REQUIRED We have on record that your facility has received the product(s) subject to this recall. DePuy Synthes asks that you review your inventory and immediately remove the affected lots from your stock. Please take the following actions: If you DO have any of the identified devices, please take the following steps: Call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number. Complete the Verification Section (page 3 of this letter) by checking the appropriate box indicating affected product has been located. Also, please indicate the number of devices found and note the Return Authorization Number. Please include your name, title, address, telephone number and signature in the spaces provided. Return the Verification Section (page 3 of this letter) with the product to Credit/Returns, DePuy Synthes, Send a copy of the completed Verification Section to DePuy Synthes by: Fax: 888-670-4162 or Scan/email: [email protected] Note: If the Verification Section is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual on page 3 of the notification. If you DO NOT have the identified product, please take the following steps: Complete the attached Verification Section (page 3 of this letter) by checking the appropriate box indicating that no affected product has been located. Please include your name, title, address, telephone number and signature in the spaces provided. This return documentation acknowledges your receipt of medical device removal information. Return the documents to DePuy Synthes by: Fax: 888-670-4162 or Scan/email: [email protected] Note: If the Verification Section is answered on
Quantity in commerce
1487
Distribution
Worldwide Distribution - US Distribution including Puerto Rico and to the countries of : Canada, Switzerland and New Zealand.
Product codes and lots
all lots of part numbers: 900.50, 900.51

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2425-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2425-2015Class IITerminatedVoluntary: Firm initiated