510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2373-2015Class II

Naturalyte Liquid Bicarbonate Concentrate 6.4 liter bottle A Dialysate Concentrate for Hemodialysis (liquid)

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2015-07-16 / Open, Classified / root cause: manufacturing process

Fresenius Medical Care Holdings, Inc. began a recall of Naturalyte Liquid Bicarbonate Concentrate 6.4 liter bottle A Dialysate Concentrate for Hemodialysis (liquid) Product Code: 08-4000-LB on . FDA classified it as Class II and gives the reason as "Expansion of recall to lots manufactured in 2015 due to the potential for bacterial contamination". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2373-2015
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Naturalyte Liquid Bicarbonate Concentrate 6.4 liter bottle A Dialysate Concentrate for Hemodialysis (liquid) Product Code: 08-4000-LB
Product code
KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
Reason for recall
Expansion of recall to lots manufactured in 2015 due to the potential for bacterial contamination.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Fresenius Medical NA issued an Urgent Expanded Medical Device Recall letter on 7/16/15 to clinics and patients. The letter identified the affected product, explained the reason for the expansion, and instructed users to immediately examine their stock to determine whether there was any Naturalyte Liquid Bicarbonate Concentrate beginning with 15AMLB, 15BMLB, 15CMLB. Users are to discontinue use immediately if any product of these lots is found, . All units are to be placed in a secure, segregated area. If affected product was on the machine prior to patient treatment, a [Heat Disinfect] program should be performed. The attached Reply Form should be completed and returned. Contact your FMCNA Customer Service Team at 1-800-323-5188 for instructions on how to return the recalled product.
Quantity in commerce
2,264,028
Distribution
Nationwide Distribution
Product codes and lots
Lots begin with: 15AMLB, 15BMLB, 15CMLB.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2373-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2373-2015Class IIOngoingVoluntary: Firm initiated