Z-2741-2015Class II

Dimension CREA (Creatinine) and Dimension Vista CREA (Creatinine); The CREA method used on the Dimension...

Siemens Healthcare Diagnostics Inc. / initiated 2015-07-15 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension CREA (Creatinine) and Dimension Vista CREA (Creatinine); The CREA method used on the Dimension clinical chemistry system is an in vitro diagnostic... on . FDA classified it as Class II and gives the reason as "Siemens has confirmed that the Dimension CREA and Dimension Vista CREA may produce falsely depressed results up to - 12.7% and -22% respectively, for a creatinine concentration of 1.5 mg/dL when hemolysis exceeds 200...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2741-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension CREA (Creatinine) and Dimension Vista CREA (Creatinine); The CREA method used on the Dimension clinical chemistry system is an in vitro diagnostic test intended for the quantitative determination of creatinine in human serum, plasma and urine.
Product code
JFY Enzymatic Method, Creatinine
Reason for recall
Siemens has confirmed that the Dimension CREA and Dimension Vista CREA may produce falsely depressed results up to - 12.7% and -22% respectively, for a creatinine concentration of 1.5 mg/dL when hemolysis exceeds 200 mg/dL and 300 mg/dL respectively.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens sent an Urgent Medical Device Correction letter, dated July 2015, was provided to all Dimension CREA and Dimension Vista CREA customers who received the products to notify them of the issue, the potential risk to health, and actions to be taken by customers. If you have any questions please contact your Siemens Customer Care Center Technical Solutions at (800) 441-9250, or your local Siemens Technical Support Representative.
Quantity in commerce
Total number Dimension Vista - 32,165, Dimension - 238,634
Distribution
Worldwide Distribution - US (nationwide) and to the countries of : Afghanistan, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia Herzegovinia, Bulgaria, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Ghana, Great Britain, Greece, Guatemala, Hungary, Iraq, Ireland, Honduras, Hong Kong, Iceland, Indonesia, India, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Libya, Macedonia, Malaysia, Mali, Mexico, Myanmar, Morocco, Netherlands, Nepal, Nicaragua, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkmenistan, United Arab Emirates, Uruguay, Venezuela, Vietnam and Yemen.
Product codes and lots
Dimension CREA Lot Numbers:  DB5195 FA5211 EA5233 DA5258 FA5275 BA5293 EA5315 EA5316 GA5342 FA5351 FA5352  FA6006 BA6026 BA6027 EB6047 GB6076 GA6097 BA6111 BA6112 FB6139 with expiration dates: 2015-07-14 2015-07-30 2015-08-21 2015-09-15 2015-10-02 2015-10-20 2015-11-11 2015-11-12 2015-12-08 2015-12-17 2015-12-18 2016-01-06 2016-01-26 2016-01-27 2016-02-16 2016-03-16 2016-04-06 2016-04-20 2016-04-21 2016-05-20  Dimension Vista CREA Lot Numbers: 14177BB 14212BD 14234BH 14261BC 14280BA 14301BB 14321AC 14346AC 15012AA 15042BA 15054AB 15076BA 15091BA 15125BB 15125BA 15141AA with expiration dates: 2015-06-26 2015-07-31 2015-08-22 2015-09-18 2015-10-07 2015-10-28 2015-11-17 2015-12-12 2016-01-12 2016-02-11 2016-02-23 2016-03-16 2016-03-31 2016-05-04 2016-05-04 2016-05-22

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2741-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2741-2015Class IITerminatedVoluntary: Firm initiated