Z-0766-2016Class II

Philips Healthcare TraumaDiagnost; Codes: 72001, 720011, 720013, 720020; A radiological examination unit for...

Philips North America LLC / initiated 2015-07-22 / Terminated / root cause: manufacturing process

Philips Electronics North America Corporation began a recall of Philips Healthcare TraumaDiagnost; Codes: 72001, 720011, 720013, 720020; A radiological examination unit for emergency work and routine skeletal diagnosis on . FDA classified it as Class II and gives the reason as "If arm is not fastened securely it could fall down". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0766-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Healthcare TraumaDiagnost; Codes: 72001, 720011, 720013, 720020; A radiological examination unit for emergency work and routine skeletal diagnosis.
Product code
KPR System, X-Ray, Stationary
Reason for recall
If arm is not fastened securely it could fall down.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Consignees were notified of the recall via letter sent certified return receipt on 7/22/2015. The letter informs customers that the Instruction for Use (IFU) requires the operator not to use the system and call for service, if a mechanical defect or malfunction is suspected. In this case, it is recommended not to position the Ceiling suspension with the Trauma arm directly above a patient as long as this Field Change Order is not yet installed. Philips plans to repair the affected systems. A Philips Service Engineer will contact customers as soon as the Field Action Kit is ready to be implemented. For questions, contact your local representative or use the Technical Support Line at 1-800-722-9377. When communicating with Philips regarding this issue, please reference Field Change Order 72000014.
Quantity in commerce
399 distributed, 262 are subject to this event
Distribution
Distributed in the states of NY, WI, MA, AL, FL, GA, RI, CA, KS, and TX, and the countries of Canada, Australia, Austria, Belgium, France, Germany, Ireland, Italy, Netherlands, New Zealand, Portugal, South Africa, Spain, and Switzerland.
Product codes and lots
All TraumaDiagnost products that were delivered before the week of July 2, 2007.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0766-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0766-2016Class IITerminatedVoluntary: Firm initiated