Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2463-2015Class II

HardyCHROM MRSA Cat no: G307 Lot no: H15155 Expires: 2015-08-13 Container Type: 15x100mm monoplate Packaged...

Hardy Diagnostics / initiated 2015-07-09 / Terminated

Hardy Diagnostics began a recall of HardyCHROM MRSA Cat no: G307 Lot no: H15155 Expires: Container Type: 15x100mm monoplate Packaged: 10 plates/sleeve Storage: 2-8C on receipt. Light... on . FDA classified it as Class II and gives the reason as "Hardy Diagnostics is recalling HardyCHROm MRSA due to a report of the potential for breakthrough of the methicillin-sensitive Staphylococcus aureus (MSSA) QC strain, ATCC 29213 on the identified lot". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2463-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hardy Diagnostics
Product
HardyCHROM MRSA Cat no: G307 Lot no: H15155 Expires: 2015-08-13 Container Type: 15x100mm monoplate Packaged: 10 plates/sleeve Storage: 2-8C on receipt. Light Sensitive. In vitro diagnostic www.HardyDiagnostics.com
Product code
JSO Culture Media, Antimicrobial Susceptibility Test, Excluding Mueller Hinton Agar
Reason for recall
Hardy Diagnostics is recalling HardyCHROm MRSA due to a report of the potential for breakthrough of the methicillin-sensitive Staphylococcus aureus (MSSA) QC strain, ATCC 29213 on the identified lot.
FDA root cause
Unknown/Undetermined by firm
Action
The firm initially notified customers via phone to inform them of a recall. The phone scripts asks customers if they have had any issued with the QC testing. The firm states that our of a precaution they are asking customers to discard the remaining plates of the identified lot, and that credit or replacement plates will be offered for any remaining. The firm followed-up phone calls with a written notification letter. The letter states to complete, sign, and fax or email back the enclosed form stating compliance with the above action. The letters states that is the customer could not be reached via phone and would like replacements or credit to contact the Customer Service Department at 800-266-2222, option 1. Any questions call the Technical Services Department at 800-266-2222, option 2 or email at techservice@hardy diagnostics.com.
Quantity in commerce
2,820 (282 pk/10)
Distribution
Distributed in the states of VA, PA, NY, NJ, IL, OH, and MD.
Product codes and lots
Cat no. G307 Lot no. H15155

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2463-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2463-2015Class IITerminatedVoluntary: Firm initiated