Concave Reamer Head, 16 mm; Concave Reamer Head, 18 mm; Concave Reamer Head, 20 mm; Concave Reamer Head 22...
Trilliant Surgical Ltd. began a recall of Concave Reamer Head, 16 mm; Concave Reamer Head, 18 mm; Concave Reamer Head, 20 mm; Concave Reamer Head 22 mm; Convex Reamer Head, 16 mm; Convex Reamer Head... on . FDA classified it as Class II and gives the reason as "The cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2456-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Trilliant Surgical LLC
- Product
- Concave Reamer Head, 16 mm; Concave Reamer Head, 18 mm; Concave Reamer Head, 20 mm; Concave Reamer Head 22 mm; Convex Reamer Head, 16 mm; Convex Reamer Head, 18 mm; Convex Reamer Head 20 mm; Convex Reamer Head 22 mm, used with Gridlock Plating system, for use in trauma and reconstructive procedures in small bones.
- Product code
- HTO Reamer
- Reason for recall
- The cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- The recalling firm sent out written notification of the recall on 7/3/14. The letter instructed the consignees to exchange affected instruments.
- Quantity in commerce
- 638 units
- Distribution
- Nationwide throughout the US
- Product codes and lots
- TSL000145, TSL000162, TSL000699, TSL000146, TSL000163, TSL000700, TSL00147, TSL000164, TSL000701, TSL000702, TSL000148, TSL000165, TSL000706, TSL000149, TSL000166, TSL000688, TSL000150, TSL000167, TSL000707, TSL000708
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2456-2015 | Class II | Terminated | Voluntary: Firm initiated |