Z-2453-2015Class II

BD CD56 PE Cy 7: NCAM16.2 ASR Analyte Specific Reagent, Catalog number 335809 Hematology: The CD 56 antigen...

Becton, Dickinson and Company, BD Biosciences / initiated 2015-07-17 / Terminated / root cause: manufacturing process

Becton, Dickinson and Company, BD Biosciences began a recall of BD CD56 PE Cy 7: NCAM16.2 ASR Analyte Specific Reagent, Catalog number 335809 Hematology: The CD 56 antigen promotes homophilic adhesion in neurons and may be... on . FDA classified it as Class II and gives the reason as "Three lots of CD56 PE Cy 7, Catalog number 335809, contain a low amount of HLA-DR PE antibody which could result in an unexpected staining pattern". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2453-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Product
BD CD56 PE Cy 7: NCAM16.2 ASR Analyte Specific Reagent, Catalog number 335809 Hematology: The CD 56 antigen promotes homophilic adhesion in neurons and may be involved in homophilic adhesion of NK cells.
Product code
MVU Reagents, Specific, Analyte
Reason for recall
Three lots of CD56 PE Cy 7, Catalog number 335809, contain a low amount of HLA-DR PE antibody which could result in an unexpected staining pattern.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
BD Bioscience sent an Urgent Product Recall Letter dated July 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customer Actions : 1. If you have CD56 PE-Cy7 integrated into a Laboratory Developed Test (LDT) or panel, evaluate the impact of a false positive in the PE channel the interpretation of on the results. 2. Discard all remaining stock of CD56 PE-Cy7, Catalog Number 335809, lot number 4300708, 4304864, and/or 5013643, and request replacement product. BD will replace product at no cost. 3. As a requirement of our Regulatory Tracking Process, we request that you complete the enclosed Tracking/Verification form, even if you no longer have the affected reagents listed in your inventory. Please return the Tracking/ Verification form to BD Biosciences, per the instructions (mail, fax, or email). We apologize for any inconvenience that this may have caused and are working toward a long-term solution to this issue. If you require further assistance, please contact BD Customer Support at 877.232.8995 (prompt 3) in the United States. For customers outside the US, contact your local BD Biosciences representative or distributor.
Quantity in commerce
362 units
Distribution
US Distribution and Internationally to Taiwan.
Product codes and lots
Catalog number 335809; Lot numbers 4300708, 4304864, 5013643; all with expiry dates of 31 OCT 2015

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2453-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2453-2015Class IITerminatedVoluntary: Firm initiated