Dimension Vista B2-Microglobulin Flex Reagent Cartridge (B2MIC)

Siemens Healthcare Diagnostics Inc. / initiated 2015-07-23 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista B2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative... on . FDA classified it as Class II and gives the reason as "customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC when using Flex reagent cartridge B2MIC various lots. In failure mode, there is the potential for falsely depressed...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2450-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Vista B2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B2-Microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B2-Microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.
Product code
JZG System, Test, Beta-2-Microglobulin Immunological
Reason for recall
customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC when using Flex reagent cartridge B2MIC various lots. In failure mode, there is the potential for falsely depressed beta2-Microglobulin results on the Dimension Vista system due to assay reagent contamination.
Root cause tag
manufacturing process
FDA root cause
Environmental control
Action
Siemens sent an Urgent Medical Device Correction letter dated July 2015 to all affected customers. The letter identified the affected product, problem, risk to health, and actions to be taken by customers. For questions contact your Customer Care Center - Technical Solutions at 1-800-441-9250 or your local Siemens technical support representative.
Quantity in commerce
Total 1655 units
Distribution
US Nationwide Distribution in the states of TN, AZ, CT, Fl, CA, NY and WI
Product codes and lots
Lot Number:15037MA; ExpDate: 2016-07-14 Lot # 15175MA; ExpDate: 2015-11-17; Lot # 15204MA; Expdate: 2017-01-13;  Lot #15246MA; expdate: 2017-02-26 Lot # 15267MA; expdate: 2017-02-26

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2450-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2450-2015Class IITerminatedVoluntary: Firm initiated