Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1
Philips Electronics North America Corporation began a recall of Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1 Product Usage: The DigitalDiagnost is intended for use in generating... on . FDA classified it as Class II and gives the reason as "The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the Microsoft Windows 7...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2383-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1 Product Usage: The DigitalDiagnost is intended for use in generating radiographic images of human anatomy by qualified/trained doctors or technicians.
- Product code
- MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Reason for recall
- The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the Microsoft Windows 7 operating system.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Philips Healthcare sent an Urgent Field Safety Notice letter dated June 25, 2015 to customers. The letter identified the affected product, problem and actions to be taken. Philips plans to install a software upgrade in affected systems, which will eliminate this issue. Philips Healthcare Field Service Engineers will schedule an appointment with customers to install the software update. For questions contact your local philips representative.
- Quantity in commerce
- 62
- Distribution
- US Nationwide Distribution to the states of : Arizona, Minnesota, State of Washington, Wyoming and Ohio. Internationally to Canada.
- Product codes and lots
- DigitalDiagnost
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2383-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY