CARESTREAM DRX-EVOLUTION X-Ray System
Carestream Health Inc began a recall of CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed diagnostic X-ray system for general radiographic x-ray imaging... on . FDA classified it as Class II and gives the reason as "Software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching inconsistencies which may go undetected on a radiograph". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2449-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Eastman Kodak Company
- Product
- CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed diagnostic X-ray system for general radiographic x-ray imaging including tomography. The tomography feature is not to be used for imaging pediatric patients.
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- Software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching inconsistencies which may go undetected on a radiograph.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Carestream sent an Urgent Medical Device Correction letters dated July 29, 2015 and Customer Notification Acknowledgement Forms to customers via Certified Mail. The letter identified the affected product, problem and actions to be taken. Carestream will update the software on your device to correct this issue. For questions or concerns, contact the Carestream Customer Care Center in the U.S. at 1-800-328-2910, available 7 days per week on a 24 hour basis. Outside of the U.S., please call your local Service support number.
- Quantity in commerce
- Domestic: 76 units, Foreign: 160 units
- Distribution
- Worldwide Distribution - US Nationwide and in the countries: Australia, Canada, Chile, China, India, Korea, Malaysia, New Zealand, Singapore, Switzerland, Taiwan and Thailand.
- Product codes and lots
- Software versions 5.6 and greater and 5.7 (current releases); Serial Numbers: 5029, 5569, 4067, 4754, 4131, 4100, 4096, 4076, 4569, 4570, 4588, 4670, 4767, 4879, 5266, 5393, 5408, 5442, 5572, 5573, 5555, 5556, 5563, 4146, 4679, 4702, 4703, 4704, 4801, 5090, 5154, 4787, 4069, 4074, 4344, 4085, 4058, 4084, 5470, 5158, 4086, 4099, 4073, 4665, 5241, 5303, 5304, 4639, 5350, 5352, 5429, 5508, 4696, 4763, 5182, 4291, 5037, 5554, 4192, 4117, 4172, 4207, 4225, 4239, 4278, 4301, 4310, 4328, 4333, 4338, 4366, 4384, 4421, 4427, 4442, 4443, 4447, 4465, 4499, 4523, 4534, 4538, 4539, 4692, 4822, 4861, 4875, 5128, 5131, 5174, 5553, 5579, 5580, 5597, 5628, 4320, 5309, 5042, 4807, 4808, 4809, 4810, 5493, 4266, 4711, 5327, 4125, 4143, 4097, 4156, 4282, 4287, 4289, 4258, 4742, 4964, 4784, 5050, 5551, 4233, 4644, 4082, 4113, 4196, 4683, 4309, 4315, 4064, 4256, 4987, 4624, 4585, 4882, 5362, 4004, 5377, 5418, 4300, 4308, 4627, 4638, 5138, 5251, 5360, 4947, 5620, 4985, 5403, 5495, 5560, 5544, 5545, 5621, 5622, 5451, 5282, 5645, 4724, 4298, 4269, 4756, 4418, 5273, 5178, 5179, 5443, 5197, 5389, 5584, 5473, 4899, 5334, 5431, 4990 --- NOTE: Additional serial numbers at foreign consignees to be determined upon reporting part usage.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2449-2015 | Class II | Terminated | Voluntary: Firm initiated |