Z-2733-2015Class II

Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of...

Siemens Healthcare Diagnostics Inc. / initiated 2015-07-28 / Terminated

Beckman Coulter Inc. began a recall of Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma... on . FDA classified it as Class II and gives the reason as "Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2733-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems.
Product code
MSW System, Test, Thyroglobulin
Reason for recall
Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL.
FDA root cause
Under Investigation by firm
Action
On 07/28/2015 the firm sent out customer notification letters via first class mail and e-mail. The firm's letter contained the following instructions: 1. Discard all Access Thyroglobulin reagent pack lots listed in the letter. 2. Evaluate Access Thyroglobulin results <10 ng/mL when using the reagent pack lot numbers listed to determine if controls and/or patients were higher than expected or did not match the clinical status of the patient. 3. At the discretion of the Laboratory Director, notify clinicians that it is possible the Access Thyroglobulin sample results reported by the laboratory were affected by this issue. The firm's notification letter requests that the information be shared with laboratory staff. If any affected products have been forwarded provide a copy of the letter to them. The firm requests that customers complete and return the Response Form within 10 days. Any questions contact the Customer Care Center: 1. From website at http://www.beckmancoulter.com 2. By phone: call 1-800-854-3633 in the United States and Canada. 3. Outside the United States and Canada, contact local Beckman Coulter representative
Quantity in commerce
3,035 (U.S.) and 7,123 (International)
Distribution
Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Algeria, Australia, Austria, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Republic of, Lebanon, Lithuania, Mexico, Morocco, New Zealand, Oman, Philippines, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Vietnam and Yemen.
Product codes and lots
Catalogue number: 33860 Lot numbers: 431990 434727 430969 522896 523251  Expiration date: 10/31/16 12/31/16 08/31/16 04/30/17 05/31/17

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2733-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2733-2015Class IITerminatedVoluntary: Firm initiated