Becton Dickinson & Co. began a recall of Vancomycin Screen Agar on . FDA classified it as Class II and gives the reason as "The affected expired lots of product may have exhibited breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life. The in-date product may also exhibit breakthrough growth...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0473-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- Vancomycin Screen Agar
- Product code
- JWY Manual Antimicrobial Susceptibility Test Systems
- Reason for recall
- The affected expired lots of product may have exhibited breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life. The in-date product may also exhibit breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life.
- FDA root cause
- Under Investigation by firm
- Action
- Becton Dickinson (BD) sent US customers a Urgent Product Recall letter dated June 2015, via UPS. Customers were requested to return response forms indicating acknowledgement of the recall and quantity of remaining product that was at their facility. The acknowledgement form included with the follow up letter that is faxed back to BD will be reconciled with the customer list. BD will contact 10% of non-respondents to verify receipt of the customer letter. Any product returned to BD will be quarantined and discarded. If further assistance is needed, please contact BD Technical Services at 1-800-638-8663.
- Quantity in commerce
- 60,840
- Distribution
- Worldwide Distribution - US (nationwide) Distribution to the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI and WV., and to the countries of : Canada, Colombia, India, Mexico and Europe.
- Product codes and lots
- catalog number 222204. Lot numbers 5041659, 5064921, 5076659, 5083712, 5097504, 5105955, 5130907, 5139626, and 5147941
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0473-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0474-2016 | Class II | Terminated | Voluntary: Firm initiated |