510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0474-2016Class II

Enterococcus Screen Agar QUAD Plate

Becton Dickinson & Co / initiated 2015-06-29 / Terminated

Becton Dickinson & Co. began a recall of Enterococcus Screen Agar QUAD Plate on . FDA classified it as Class II and gives the reason as "The affected expired lots of product may have exhibited breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life. The in-date product may also exhibit breakthrough growth...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0474-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
Enterococcus Screen Agar QUAD Plate
Product code
JWY Manual Antimicrobial Susceptibility Test Systems
Reason for recall
The affected expired lots of product may have exhibited breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life. The in-date product may also exhibit breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life.
FDA root cause
Under Investigation by firm
Action
Becton Dickinson (BD) sent US customers a Urgent Product Recall letter dated June 2015, via UPS. Customers were requested to return response forms indicating acknowledgement of the recall and quantity of remaining product that was at their facility. The acknowledgement form included with the follow up letter that is faxed back to BD will be reconciled with the customer list. BD will contact 10% of non-respondents to verify receipt of the customer letter. Any product returned to BD will be quarantined and discarded. If further assistance is needed, please contact BD Technical Services at 1-800-638-8663.
Quantity in commerce
770
Distribution
Worldwide Distribution - US (nationwide) Distribution to the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI and WV., and to the countries of : Canada, Colombia, India, Mexico and Europe.
Product codes and lots
catalog number 222201.  Lot numbers 5064938, 5085954, 5105509, and 5133580

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0474-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0473-2016Class IITerminatedVoluntary: Firm initiated
Z-0474-2016Class IITerminatedVoluntary: Firm initiated